Shopping Cart

0 item(s) - 0.00$
Your shopping cart is empty!
Drop us a line
T: 514-354-2511
T: 1 800 463-4363
F: 514-354-6948
Biosafety cabinet, Class II, Type A2, NSF 49

Biosafety cabinet, Class II, Type A2, NSF 49

Product Code:

Need more Info? (514) 354-2511 - 1-800-463-4363

NSF Certified
CFDA Certified UL61010-1/CSA-C.22.2 No.61010-1 Certified
ISO 9001:2008 Quality Management System Certified


Multiple Protection Design Increases Security Level Significantly
Duplex electrical outlets with splash-proof cover.
(UL listed)
Service valve for gas, air and vacuum. Maximum working pressure 75PSI. (CSA Certified)


User-Friendly Operation and Maintenance
  • Apply Key to switch power for operator security control.
  • Ergonomic 10 degree angle window design. Reduces operation discomfort over long-term use of limbs and eye strain.
  • Work area and work tray are fabricated in one single-piece of stainless steel with rounded corners designed to reduce particulate adhesion and for easy cleaning and maintenance.
  • Patented HEPA fixation design for a more secure locking device. Can replace HEPA filters from the front of cabinet quickly and easily.
  • Pioneered single-piece stainless steel design of work tray with front grille. The front grille Arc increases the fluency of the inflow air. It can be used as an armrest and eliminates the need for additional armrest fittings.
  • Removable single-piece work tray can be easily removed for cleaning and collection of spilled liquid.


Push Switch Panel
Window alarm When the window is not at the operation height of 20 cm, it will release an audible and visual alarm automatically.
Mute switch When the alarm is activated, it can be muted temporarily while the alarm light continues to flash. If the abnormal situation has not been settled, the cabinet alarm will be activated again within a set time (5 minutes) to alert the user.
Safety interlock When the window is opened, the fan and fluorescent light will be activated automatically and the UV light will be switched off. When the window is fully closed, the UV light will be re-activated automatically. The fan and fluorescent light will be switched off.
※ UV LIGHT TIMER for extra order.


Detailed Specifications
Cabinet Type 4 feet
International Certification ● NSF 49
● EN12469
● UL61010-1/CSA-C.22.2 No.61010-1
● ISO 9001:2008
External Dimensions (WxDxH)
(without support stand 640mm)
1430 x 887 x 1650 mm
Internal Dimensions (WxDxH) 1245 x 587 x 637 mm
Cabinet Main Construction Stainless Steel#304
Air Flow 70% recirculated / 30% exhausted
Window Operation Height 20cm (8")
Ave. Inflow Velocity 0.51m/s
Ave. Downflow Velocity 0.375m/s
Exhaust Air Volume with Exhaust Canopy 16m³/min
Supply/Exhaust HEPA Filter 99.99% @0.3 micron (Class H14)
Light Intensity > 800 lux
Sound Level < 65 dBA
UV Lamp 254 nm / FL 30W*1
Fluorescent Lamp T5 / 28W*2
Optional Accessories ● IV bar with 6 hooks
● Service valves (air, gas, vacuum), CSA Certified
● Exhaust canopy
Room Height Suggested 250cm at least
Power Consumption 115V,60Hz,10A 430W


Ordering Guide - Model Number
Cabinet Type 4 feet
Touch Panel / 115V,60Hz,10A GE1430AAA-0
Push Switch Panel / 115V,60Hz,10A GE1430BAA-0