- Uses optical DO technology
- No membranes to change –
once-a-year cap replacement
- Calibrate for dissolved oxygen
using water-saturated air,
air-saturated water, zero point or a
custom point with Winkler titration
- Built-in barometer for automatic
pressure correction with option for
manual input
- Temperature calibration capability
- Data storage for 1000 data sets
with time and date stamping
- RS232 port for easy data
downloading and software updates
- IP67-rated meter housing and
backlit display
- IP68-rated RDO sensor
- Four AA batteries provide over
1000 hours of operation in
AUTO-READ™ mode
- 3-year warranty on meter and
RDO sensor body
- 1-year warranty on optical cap
|
DO
|
Concentration
Range |
0.00 to 20.0 mg/L |
| |
Resolution |
0.01, 0.1 mg/L |
| |
Relative
Accuracy |
±0.1 mg/L up to
8 mg/L; ±0.2 mg/L
from 8 to 20 mg/L |
| |
% Saturation
Range |
0.0 to 200% |
| |
Resolution |
0.1, 1% |
| |
Relative
Accuracy |
± 2% |
| |
Barometric
Pressure
Correction |
450 to 850 mm Hg,
automatic or manual |
| |
Calibration
Features |
Water saturated air,
air saturated water,
manual (Winkler) and
zero point |
| |
Sensor Type |
RDO optical |
| Temperature |
Range |
0 to 50 °C |
| |
Resolution |
0.1 °C |
| |
Relative
Accuracy |
± 0.3 °C |