Orion 4-Star Benchtop pH/ISE Meter

Benchtop Meter (Cat. No. 1115000)
Benchtop Meter Kit (Cat. No. 1115001)
• Easy to use, 3-button function
• SMART STABILITY™ and SMART AVERAGING™ functions for automatically optimized accuracy, precision and response time
• Storage of up to 10 password protected methods for easy retrieval of operation procedures
• Accepts ISFET (Ion Selective Field Effect Transistor) electrodes
• Simultaneous display of pH or ISE and temperature measurements
• Auto stirrer control

Specifications
pH
Readout Range / Resolution
-2.000 to 19.999 or 0 to 14 in aqueous solutions / 0.1,0.01,0.001
Relative Accuracy
± 0.002
SMART STABILITY™ and SMART AVERAGING™
Resolution dependent / automatic
AUTO-CAL™ and/or Manual Calibration 1 to 5 points; US/NIST, DIN, and custom buffers recognized
ISE
Range / Resolution
0 to 19900 / 1,2, or 3 signi•cant digits
Relative Accuracy
± 0.2 mV or 0.05%, whichever is greater
SMART STABILITY™ and SMART AVERAGING™
Resolution dependent / automatic
Units
M, mg/L, %, ppb, or no units
# ISE Cal points
1 to 5
Special ISE Calibration Features Linear point to point, selectable auto-blank non-linear, low concentration range
mV
Range – mV / RmV
± 1999.9
ORP
EH autocalibration to standard hydrogen electrode
Resolution
0.1
Relative Accuracy ± 0.2 mV or 0.05%, whichever is greater
Temperature
Range / Resolution
-5 to 105 °C / 0.1 up to 99.9 °C, 1.0 over 99.9 °C
Relative Accuracy
± 0.1 °C
Temperature Compensation
Auto/Manual
Display Custom backlit LCD, for easy viewing
Inputs/Outputs
Sensor Inputs
pH/ISE - BNC; ATC-miniDIN-Locking; Reference-Pin tip; Stirrer-Phono-jack
Solid State pH Sensor
ISFET - miniDIN - locking
Communications Bi-directional RS232 - Stereo jack, optional USB interface Cable
Special GLP Software Features
# Of Data Logging Points
200 points by time, stability or manual read with time/date stamp
Calibration Log
Last 10 Calibrations
User Method Storage
10 Saved
Advanced Self-test and Diagnostics
YES
GLP Electrode Verification
YES
Password Protection
YES
Internet Software Updates / Downloadable Methods
YES
21 CFR Part 11 Software Package
Optional
Power
Universal power adapter and battery power 4 x AA batteries
Regulatory and Safety CE, CSA, UL, TÜV, FCC Class A limits
Environmental Operating Conditions
Ambient Operating Temperature
5 to 45 °C
Relative Humidity
5 to 85% Non-Condensing
IP Rating
IP54, splashproof and dustproof
Warranty 36 months (from date of purchase)
Above data generated under optimal/controlled circumstances.
Note: ISFET capability available in Fall 2005

 

pH/ISE Benchtop Meter Kit (1115001) Includes:
Cat. No.
Description
1115000
4-Star pH/ISE benchtop meter
8102BNU
ROSS Ultra® glass pH electrode
917005
A series meter epoxy ATC probe
810001
ROSS™ pH electrode storage solution, 475 mL
1110001
Electrode stand
910104
pH 4.01 buffer, 475 mL
910107
pH 7.00 buffer, 475 mL
910110 pH 10.01 buffer, 475 mL