| pH-mètre de table |
| modèle 4-STAR de Thermo (Orion) |
| Portable Meter (Cat. No. 1215000) Portable Meter Kit (Cat. No. 1215001) Durable, waterproof, IP67 meter stands up to the most demanding field applications while delivering over 2,000 hours of continuous operation. Ergonomic design for comfortable handling Accepts ISFET (Ion Selective Field Effect Transistor) electrodes SMART STABILITY and SMART AVERAGING functions for automatically optimized accuracy, precision and response time Storage of up to 10 password protected methods for easy retrieval of operation procedures Simultaneous display of pH or ISE and temperature measurements |
| Specifications | |
| pH | |
| Readout Range / Resolution | -2.000 to 19.999 or 0 to 14 in aqueous solutions / 0.1,0.01,0.001 |
| Relative Accuracy | ± 0.002 |
| SMART STABILITY and SMART AVERAGING | Resolution dependent / automatic |
| AUTO-CAL and/or Manual Calibration | 1 to 5 points; US/NIST, DIN, and custom buffers recognized |
| ISE | |
| Range / Resolution | 0 to 19900 / 1,2, or 3 signicant digits |
| Relative Accuracy | ± 0.2 mV or 0.05%, whichever is greater |
| SMART STABILITY and SMART AVERAGING | Resolution dependent / automatic |
| Units | M, mg/L, %, ppb, or no units |
| # ISE Cal points | 1 to 5 |
| Special ISE Calibration Features | Linear point to point, selectable auto-blank non-linear, low concentration range |
| mV | |
| Range mV / RmV | ± 1999.9 |
| ORP | EH autocalibration to standard hydrogen electrode |
| Resolution | 0.1 |
| Relative Accuracy | ± 0.2 mV or 0.05%, whichever is greater |
| Temperature | |
| Range / Resolution | -5 to 105 °C / 0.1 up to 99.9 °C, 1.0 over 99.9 °C |
| Relative Accuracy | ± 0.1 °C |
| Temperature Compensation | Auto/Manual |
| Display | Custom backlit LCD, for easy viewing |
| Inputs/Outputs | |
| Sensor Inputs | pH/ISE - BNC; ATC-miniDIN-Locking; Reference-Pin tip (all waterproof) |
| Solid State pH Sensor | ISFET - miniDIN - locking |
| Communications | Waterproof bi-directional RS232 - Stereo jack, optional USB interface Cable |
| Special GLP Software Features | |
| # Of Data Logging Points | 200 points by time, stability or manual read with time/date stamp |
| Calibration Log | Last 10 Calibrations |
| User Method Storage | 10 Saved |
| Advanced Self-test and Diagnostics | YES |
| GLP Electrode Verification | YES |
| Password Protection | YES |
| Internet Software Updates / Downloadable Methods | YES |
| 21 CFR Part 11 Software Package | Optional |
| Power | 4 x AA batteries, powerless date protection |
| Regulatory and Safety | CE, CSA, UL, TÜV, FCC Class A limits |
| Environmental Operating Conditions | |
| Ambient Operating Temperature | 5 to 45 °C |
| Relative Humidity | 5 to 85% Non-Condensing |
| IP Rating | IP67, waterproof and dustproof |
| Warranty | 36 months (from date of purchase) |
| Above data generated under optimal/controlled circumstances. | |
| Note: ISFET capability available in Fall 2005 | |
| pH/ISE Benchtop Meter Kit (1115001) Includes: | |
| Cat. No. | Description |
| 1215000 | 4-Star pH/ISE portable meter |
| 9107BNMD | Low maintenance Triode pH/ATC electrode |
| 910060 | pH storage solution, 5x60 mL |
| 1210004 | Hard field case |
| 910410 | pH 4.01 buffer, 10 pack |
| 910710 | pH 7.00 buffer, 10 pack |
| 911010 | pH 10.01 buffer, 10 pack |