| pH/RDO/Conductivity Portable Meter |
| Model 5-Star |
![]() |
![]() |
| Full-featured Portable Multi-parameter System with RDO (Rugged Dissolved Oxygen) Technology |
|
| Specifications |
DO |
Concentration
Range |
0.00 to 20.0 mg/L |
Resolution |
0.01, 0.1 mg/L |
|
Relative
Accuracy |
±0.1 mg/L up to
8 mg/L; ±0.2 mg/L
from 8 to 20 mg/L |
|
% Saturation
Range |
0.0 to 200% |
|
Resolution |
0.1, 1% |
|
Relative
Accuracy |
± 2% |
|
Barometric Pressure
Correction |
450 to 850 mm Hg,
automatic or manual |
|
Calibration
Features |
Water saturated air,
air saturated water,
manual (Winkler) and
zero point |
|
Sensor Type |
RDO optical |
|
pH |
Range |
-2.000 to 19.999 |
Resolution |
0.1, 0.01, 0.001 |
|
Relative
Accuracy |
± 0.002 |
|
mV, RmV,
ORP |
Range |
± 1999.9 mV |
Resolution |
0.1 mV |
|
Relative
Accuracy |
± 0.2 mV or 0.05%
of reading, whichever
is greater |
|
Conductivity |
Range |
0.000 to 3000 mS/cm,
auto resolution, cell
constant dependent |
Resolution |
4 significant digits down
to 0.001 μS/cm, cell
constant dependent |
|
Relative Accuracy |
0.5 % ± 1 digit or
0.01 μS/cm, whichever
is greater |
|
Cell Constants |
0.001 to 199.9 cm–1 |
|
Resistivity |
Range |
0.0001 to 100 Megohm |
Resolution |
Automatic |
|
Relative Accuracy |
0.5 % ± 1 digit |
|
Salinity |
Range |
0.1 to 80 ppt NaCl
equivalent, 0.1 to 42 ppt
practical salinity |
| Resolution |
0.1 ppt |
|
Relative Accuracy |
± 0.1 ± 1 digit |
|
TDS |
Range |
0 to 19999 mg/L |
Resolution |
1 mg/L |
|
| Relative Accuracy |
± 0.5% ± 1 digit |
|
Temperature (with RDO
sensor) |
Range |
0 to 50 °C |
Resolution |
0.1 °C |
|
Relative
Accuracy |
± 0.3 °C |
|
Temperature (when RDO
sensor is not
used) |
Range |
-5 to 105 °C |
Resolution |
0.1 up to 99.9 °C,
1.0 over 99.9 °C |
|
Relative
Accuracy |
±0.1 °C |