- Conforms to safety requirements IEC 1010-1; 1010-2-D
- 4 place swing out rotor
- Static, near vertical bucket positioning
- Specifically designed for testing oils
- Heated chamber, ambient to 80°C
- Membrane ‘touch’ control panel
- Pre-heat facility to ensure bowl is at test temperature
prior to test
- Universal buckets for 12.5ml, 6 and 8 inch glassware
- Optional ‘Setafuge’ Hazard Reduction System
- Microprocessor control with 99 memories
- Optional trolley
|
A four place swing out rotor and universal bucket assembly accommodates
low cost adaptors and cushions suitable for borosilicate glass centrifuge
tubes of the types specified for oil testing: 6 or 8 inch conical, 8 inch trace
sediment, 8 inch pear-shaped or 12.5ml. To reduce errors in measurement
of tube content, the rotor design allows tubes to stand in a near upright
(vertical) position at rest.
Speed, temperature, and timing are microprocessor controlled and run
parameters are entered via the panel controls and digital displays. Speed
can be specified and measured in either RPM (rotational speed) or RCF
(relative centrifugal force) mode and 99 run profiles can be stored in
memory. The run duration can be set between 10 seconds and 100 hours in
10 second and one-minute increments, or set for continuous operation.
Acceleration and braking time are user adjustable.
A controlled isothermal band heater holds the sample temperature
between 30 and 80°C, and is programmed to compensate for friction
heating effects during operation thus ensuring correct sample
temperatures throughout the test. A brushless asynchronous motor ensures
high reliability with low maintenance and the centrifuge has in-built error
messaging diagnostics.
SAFETY features include a motor driven lid latch with a safety interlock to
prevent centrifuge operation while the lid is unlocked, imbalance
detection, and over-temperature protection for both the motor and the
chamber. The centrifuge is almost silent in operation with measured noise
levels below 55 dB (A). Two types of trolleys are available as optional
accessories to allow easy movement of the instrument. The centrifuge
meets the safety requirements of BS 4402; DIN 58970; IEC 1010-1; IEC
1010-2-D; pr EN 61 010-1 & 61 010-1; BGR 500; Chapter 2.11; Part 3; UVV
VBG4; and VDE 0871. |