This instrument is a portable 31/2 digit., compact-sized digital thermometer designed to use external K-type thermocouple as temperature sensor. Temperature indication follows National Bureau of Standards and IEC584 temperature/voltage tables for K-type thermocouples. One K-type thermocouples is supplied with the thermometer


Features:
  1. Type K thermometer with single inputs direct or differential measurements to 0.1°
  2. Electronic Offset function allows compensation of thermocouple errors to maximize overall accuracy.
  3. Readout in ºC,ºF, or Kelvin (K).
  4. Auto Power Off mode (Sleep mode) increases battery life.
  5. Max hold and Data hold
  6. Wide variety of interchangeable Thermocouple Probes

Accuracy:
Accuracy is specified for operating temperatures over the range of 18ºCto 28ºC(64ºF to 82ºF), for l year, not including thermocouple error.

±2ºC
±4ºF
±5K
±(0.5% rdg+1ºC)
±(0.8% rdg+1ºC)
±(0.5% rdg+2ºF)
±(1.0% rdg+2K)

-50ºC to 0ºC
-58ºF to 32ºF
223K to 273K
0ºC to 1000ºC
1000ºC to 1300ºC
32ºFto 2000ºF
273Kto 2000K

Specifications:

Measurement Range:
-50ºC to 1300ºC, -58ºF to 2000ºF, 223K to 2000K
Resolution: 1ºC or 1ºF, 0.1ºC or 0.1ºF,1 K
Temperature Coefficient:
0.1 times the applicable accuracy specification per ºC from 0ºC to 18ºC and 28ºC to 50ºC (32ºF to 64ºF and 82ºF to 122ºF).
Input protection:
60V dc or 24V rms ac maximum input voltage on any combination of input pins.
Reading Rate: 2.5 times per second.
Input Connector:
Accepts standard miniature thermocouple connectors (flat blades spaced 7.9mm, center to center).
ENVIRONMENTAL
Ambient Operating Range:
0ºC to 50ºC (32ºF to 122ºF)
Storage Temperature:
-20ºC to 60ºC (-4ºF to 140ºF)
Relative Humidity:
0% to 80% (0ºC to 35ºC) (32ºF to 95ºF)
0% to 70% (35ºC to 50ºC) (95ºF to 122ºF)
GENERAL
Display:
31/2 digit liquid crystal display (LCD) with maximum reading of 1999
Battery:
Standard 9V battery
Supplied Probe:
4 foot type"K"thermocouple bead probe (Teflon tape insulated).
Maximum insulation temperature 260ºC(500ºF)
Probe accuracy:
±2.2ºC or ±0.75% of reading (Whichever is greater) from 0°to 800ºC
Dimensions:
162mm (H)◊76mm(W)◊38.5mm(D)
Weight: 210g