Thermo Scientific Orion Star A112
Conductivity Benchtop Meters
combine simplicity with accuracy.
A large LCD displays conductivity or
TDS readings along with temperature.
Icons provide quick updates on battery
life and completed calibration status.
Simple button layout and onscreen
messages help with calibration and
setup menu choices. |
- Large, informative screen clearly
shows the main information,
including conductivity or TDS
reading, temperature in ºC or ºF,
meter mode and power source
- Don’t miss a reading – AUTOREAD ™ locks in the stable reading
on your screen and ready indicator
alerts when readings are stable
- Selectable cell constant
- Selectable reading reference
temperatures of 20 °C or 25 °C
for an accurate result
- Included electrode arm and
newly-designed holder make
it easier to maintain and place
probes into samples
- Plug the meter into almost any
power outlet with the included
universal power adapter, or use
four AA batteries (sold separately)
to run the meter on DC power
- Non-volatile memory holds up to
50 data points
- The IP54-rated housing can handle
splashes and can be wall-mounted
if table space is a problem
- 3 year meter replacement warranty
|
| Conductivity |
| Range |
0.1 μS/cm to 200 mS/cm |
| Resolution |
0.01 μS minimum;
3 significant figures minimum |
| Relative Accuracy |
0.5 % reading ±1 digit |
| Reference
Temperature |
20, 25 °C (default) |
| Temperature
Compensation |
Linear (0.0 to 10.0 %/ºC) |
| Compatible
Cell Constants |
0.001 to 10 |
| Number of
Calibration Points |
1 |
| TDS |
| Range |
1 to 19,999 mg/L |
| Resolution |
4 significant digits |
| Relative Accuracy |
0.5 % reading ±1 digit |
| TDS Factor Range |
Linear 0.01 to 10.00,
default 0.49 |
| Temperature |
| Range |
-5 to 105 °C, 23 to 221 ºF |
| Resolution |
0.1 |
| Relative Accuracy |
±0.1 |
| Temperature Offset
Calibration |
Yes |
| Datalogging |
| Number of Points |
50 |
| Log Function |
Manual, Automatic with AUTO-READ™ |
| Log Edit |
Delete last reading or all |
| Datalogging |
| Number of Points |
50 |
| Log Function |
Manual, Automatic
with AUTO-READ™ |
| Log Edit |
Delete last reading or all |
| Inputs |
| Conductivity with
Temperature Probe |
8-pin mini-DIN |
| Power |
| AC Adapter |
Included– universal, 100-240 VAC |
| Battery Power |
Optional– 4 AAs |
| Battery Life |
2000 hrs |
| CAT. No. |
Description |
| Meters Only |
| STARA1120 |
Orion STAR A112 Conductivity Benchtop Meter
- Electrode arm with redesigned holder
- Universal power adapter |
| STARA1125 |
Orion STAR A112 Conductivity Benchtop Meter Kit
- 011050MD orion conductivity cell, K=1.0
- 011007 conductivity standard, 1413 μS, 5x60 mL
- Electrode arm with redesigned holder
- Universal power adapter |
| Accessories |
| STARA-BEA |
Benchtop electrode arm for Orion Star
A-series meters |
| STARA-HB |
Freestanding base for use with Orion Star A-series
benchtop electrode arm |
| 011050MD |
Orion 2-electrode conductivity cell, K=1.0 |
| 013016MD |
Orion 2-electrode conductivity cell with
flow cell, K=0.1 |
| 011008 |
Orion conductivity standard, 100 μS/cm,
5x60 mL bottles |
| 011007 |
Orion conductivity standard, 1413 μS/cm,
5x60 mL bottles |
| 011006 |
Orion conductivity standard, 12.9 mS/cm,
5x60 mL bottles |
| 1010001 |
Orion conductivity calibration resistor kit |