GENEQ > Biotechnologie > Bilirubine – Analyseur
| Analyseur de bilirubine |
| Modèle BR2 Stat-Analyzer |
The
Advanced Model BR2 Bilirubin Stat-Analyzer® provides a rapid, accurate
determination of total and direct bilirubin in neonates. And because it's
so compact and easy to use, it's appropriate for virtually any laboratory. Advanced Instruments products are praised for their reliability, ease-of-use, and small sample size. The world's leading manufacturer of freezing-point osmometers for clinical and research applications, Advanced Instruments, Inc. designs and manufactures innovative, value-based solutions for manufacturing, quality control, microbiology and research laboratories. |
| The Model BR2 is the most convenient method - it's the only instrument that measures total and direct bilirubin from a single 30 µL sample. |
|
| Clinical Implications |
|
Bilirubin is a by-product of the breakdown of hemoglobin. When hemoglobin is metabolized, insoluble bilirubin is produced. The insoluble (indirect) bilirubin is broken down into watersoluble (conjugated or direct) form by the liver. Direct bilirubin is then excreted. Hyperbilirubinernia (high levels of insoluble bilirubin) results in jaundice. Most neonates have some degree of jaundice shortly after birth, because the liver is not fully developed. However, prolonged hyperbilirubinernia can indicate a serious condition. It can lead to kernicterus which may result in mental retardation, serious neurologic damage, or death. |
| Principles of Operation |
The BR2 determines bilirubin levels
through spectrophotometric analysis. A beam of light passes through the
serum or plasma sample and is split into two beams. One detector measures
absorbance at 454 nm and another detector measures absorbance at 540 nm.
The two measurements are compared and total bilirubin is determined. Direct bilirubin is determined by the method of EvelynMalloy, after Henry. This same sample is acidified and a diazo reagent is added. A color reaction with the direct bilirubin is induced, and direct bilirubin concentration is measured. |
| Advanced Bilirubin Stat-Analyzer Specifications | |
| Sample Size | 30µL |
| Sample Capacity | Single sample |
| Test Time | Total: Instant Direct: 2 minutes |
| Units | mg/dL bilirubin (also available in µ mol/L) |
| Range | 0 to 30 mg/dL |
| Linearity | Within 0.2 mg/dL from straight-line |
| Resolution | 0. 1 mg/dL |
| Repeatability | ±O.2 mg/dL (1 S.D.) |
| Power Requirement | 100 to 130 V (60 Hz) or 200 to 250 V (50 Hz) |
| Power Consumption | 20 W |
| Dimensions | 7"H x 11"W x 13"D (18 cm H x 28 cm W x 33 cm D) |
| Net Weight | 9 lbs. (4 kg) |
| Shipping Weight | 12 lbs. (5.5 kg) |
| Storage Temperature | -40 to +160ºF (40 to +70ºC) |
| Operating Temperature | 59 to 90ºF (15 to 32'Q |
| Operating Humidity | 5 to 80% R.H., non-condensing |
| Warranty | One-year limited warranty on workmanship and all parts except glass, plastic and parts warranted by their makers. |