GENEQ > Environment > Air Samplers
| Air Flow Calibrators and Adapters |
Low & Medium & Hi Volume Air Flow Calibrators, Analog & Digital Displays
| AIR FLOW MEASUREMENT DEVICES |
Our steel constructed,two-component
polyurethane painted Air Flow Calibrator (AFC) housings incorporate a
precision machined venturi tube to create a pressure differential across
a fixed orifice.By varying the air flow velocities,the pressure differentials
are measured &recorded against a laminar flow element whose calibration
is certified and traceable to the National Institute of Standards and
Technology (N.I.S.T.).The AFC s overall internal simplicity of operation
reduces the frequency of recalibration to once or twice per year depending
on (individual)regulatory protocol.AFC-XX series air flow calibrators
are intended to be used open to air. |
Dual Range AFC |
| AFC-SERIES FLOW RANGES |
| By varying the orifice size and the chosen Magnehelic ® ((Standard AFC) or differential pressure sensor/transmitter (AFC-DIGITAL), We can fabricate calibrators in almost any desired flow range.Unlike other manufacturers,we do not use pre-silk screened dial faces in their analog AFC units. GENEQ individually hand marks each AFC dial face. Flow vs.Pressure Differential produces a logarithmic reading on a Magnehelic ® gage. The spacing between the numbers on the lower end of the scale will be close together and further apart on the upper end of the scale. For the greatest visual accuracy, GENEQ suggests, ordering a unit with the lowest upper range as practical (i.e.:don t select an AFC-8 if your pump only draws a maximum of 4 CFM). |
|
Analog style
AFC |
Analog style AFC units house a Magnehelic
® differential pressure gage, whose relative volumetric flow rate
units are transferred onto a direct reading scale.A reflective tape on
the scale,behind the analog needle,reduces parallax error. Overall accuracy
on analog AFC units is better than ±5%. |
|
Digital style
AFC |
Digital style AFC units house an LED
display which employs advanced technology for stable,drift free,readout
while incorporating features that provide flexibility now and in the future
with Plug-in option cards available for alarm contacts,analog output,and
serial communication.The AFC-DIGITAL s Rate/Total dis- play indicates
air flow rate in Actual Cubic Feet per Minute (ACFM) and displays total
volume sampled in Actual Cubic Feet (ACF).Metric calibration is also available.Accuracy:+/-1%full
scale at calibration points.(15 points per standardcalibration).Contact
GENEQ for Analog to Digital Conversion kits. |
|
Dual Range ACF |
If you require visual
accuracy over a large flow range,or wish to calibrate a number of air
flow units consisting of more than one flow range,select GENEQs
Dual Range AFC.The Dual Range unit consists of two pressure differential
gauges installed in a single cabinet,that precisely measures both the
low and high rate of air through a single venturi. |
| HI-VOLUME AIR FLOW CALIBRATOR, HFC-SERIES |
| The HFC-Series,HI-Volume Air Flow Calibrators,have eliminated the need for cumbersome orifice plates and water manometers.The HFC-XXC series units utilize a precision machined Venturi tube coupled with a pressure differential gauge giving a direct reading in the volumetric units of your choice (i.e.SCFM @stp).The unit is calibrated against an in-line N.I.S.T.traceable laminar flow element.The primary calibrator meets the requirements of MIL Std.45662A.HFC-XX series units are intended to be used open to air. The direct meter read-out will indicate the flow in standard CFM,LPM,or CMM at standard conditions (29.92"of Hg and 70 º F)). Given actual sampling tempera- ture &barometric pressure during calibration,a technician can convert actual flow readings (i.e.ACFM)to standard units (i.e.SCFM) by making a simple calculation using look-up correction factors from tables given in the operating manual. Depending on orientation at calibration (vertical vs.horizontal), GENEQ offers two HFC models:The HFC-XX is intended for high flow calibrations in the verticalplane,such as seen with the HVP style high volume air samplers..Whereas HFC-SIDE-XX is intended for the calibration of low to high flow rates in the horizontal plane,such as seen with the line of portable continuous duty and grab air samplers. The HFC-XX series units have a standard 1 1ž2"male straight pipe thread (MSPT) which can be screw into most low to medium volume air sampling pumps (ie:CF-1001 or CF-901 ).On higher volumetric flow samplers,a user must incorporate an adapter plate.With the use of FHA-4CF 4"diameter adapter disc,an operator is able to calibrate a unit with a 4"diameter filter holder assembly. The FHA- 810CF adapter plate is used to calibrate units with an 8"x 10"paper holder. |
| Ordering Information |
Replace XX with the uppermost
desired flow rate..If a dual scale reading is required (i.e.both CFM &LPM
displayed on one scale) replace C with DSC (e.g.order:HFC-50DSC
for 10-50 CFM &280-1500 LPM marked scale) High Volume,Single Range
Air Flow Calibrators ranges be- tween 10-50 CFM (280-1500 LPM) |
![]() |
| AIR FLOW MEASURING UNIT - V-FLO SERIES |
The V-FLO-XX series units
incorporate a flowmeter and a precision- machined venturi tube to create
a pressure differential across a fixed orifice which floats the indicating
ball in the flowmeter.By varying the flow velocities,the pressure differentials
are measured &recorded against a laminar flow element whose calibration
is certified and traceable to the National Institute of Standards and
Technology (N.I.S.T.). The recorded volumetric flow units are then marked
directly on the attached flowmeter scale. V-FLO models are accurate within
±5%of full scale reading. The standard V-FLO-XX body is six inches
long and threaded on bothends with 3/8"female NPT. The complete unit
includes one male (outlet)&one female (inlet)3/8"quick disconnect
fitting.The V-FLO XX,venturi flow meters are intended for in-line use.
Ordering Information for Venturi Flow meters:On custom units replace XX
with the uppermost desired flow rate. If a dual scale reading is
required (i.e.both CFM &LPMdisplayed on one scale)add the letters
DSC to the end of the model number ((e.g.order:V-FLO-5-DSC
for 1-5 CFM &28-140 LPM marked scale). Note:Larger Flow meters with
expanded scales are available. |
![]() |
| CALIBRATOR ADAPTER FITTINGS |
|
Air Flow Calibrators are generally used to verify the displayed sample flow rate accuracy of complete air sampling systems.Along with the flow measuring device to be calibrated,it is recommended that a Complete system for calibration should also includes the actual filter holder and filter media in order to duplicate realistic pressure drops experienced in a true sampling environment.The following adapters allow a technician to connect an Air Flow Calibrator at the end of (open to air)an In-Line High,,Medium or Low Volume Air Sampling calibration set up. Standard Calibration set-up,from right to left,(as shown in picture)is as follows:1.) Open to Air, 2.) Air Flow Calibrator, 3.) Calibrator Adapter Fitting, 4.) Filter Holder &Media, 5.) Flow meter to be Calibrated, 6.) Flow Control Valve, and finally 7.) Vacuum Pump. |