GENEQ > Environnement > Air-Qualité
| Moniteur de la qualité de l'air extérieur |
| Modèle AQM 60 |
| Features |
| • High quality data • Weatherproof FRP enclosure • Portable & compact system • Singular or multi-gas sensors • Temperature & humidity option • Wind speed & direction option • Wall or pole mounting options • Real time or data log monitoring • Modular components for flexible configuration & integration • Interchangeable sensor modules for easy maintenance • GSM, RF modem, 802.11 wireless, Bluetooth®, direct connection |
| Applications |
| • Urban air quality monitoring • Industrial air quality monitoring • Remote air quality data collection • Perimeter monitoring (landfill & wastewater sites) |
| Interior View (sensor modules removed) | |
![]() |
![]() |
| General Specifications | ||
| GSS Gas Sensor Modules (up to 6 modules) |
O3 | 0 – 0.150 ppm; 0 – 0.5 ppm |
| NO2 | 0 – 2 ppm | |
| VOC | 0 – 500 ppm | |
| CO | 0 – 500 ppm | |
| CO2 | 0 – 2000 ppm | |
| SO2 | 0 – 10 ppm | |
| H2S | 0 – 1 ppm; 0 – 10 ppm | |
| NH3 | 0 – 500 ppm | |
| See full sensor specifications for accuracy,
LDL, temp, RH etc (Other gases and ranges available on request) |
||
| Temperature & Relative Humidity Module | Range | -20 to 100°C |
| 0 to 100% RH | ||
| Power Supply Module | Input | 100 – 240 VAC |
| Output | 12 VDC Regulated | |
| Communications Module | GSM; RF modem; 802.11 wireless; Bluetooth®;
landline modem; CAT6 direct connection * |
|
| Control Module | Programmable Logic Controller (PLC) * Micro-controller (AQL software provided) Real time monitoring with Internet options PC or PLC download options (Other data storage capacities available on request) |
|
| Enclosure | Size | 393 x 342 x 220 mm; 15.5 x 13.5 x 8.7 in. |
| Weight | 9 Kg (approx.) | |
| Rating | NEMA 3R / IP42 ; UL 94-5V (flame rating) | |
| Fan | Internal with vent | |
| Colour | Machine tool grey | |
| Material | Fibreglass reinforced polyester (FRP); stainless steel lockable latches | |
| Mounting | Fixed | Direct screw mount; optional SS bracket for pole, fence or tree mounting |
| Optional Extras | Wind speed & direction sensor Solar panel power supply Internal heater * Options are user definable |
|
| Enclosure Dimensions |
![]() |
| Sensor Specifications | |||||||
| Sensor | Calibrated Range (ppm) |
Lowest Detection Limit |
Accuracy | Resolution | Reading Update |
Ambient Operating Temperature |
Relative Humidity (non-condensating) |
| NH3 | 0 - 500 | 0.5 ppm | <± 5 ppm < 100 ppm <± 10% 100 - 500 ppm |
1 ppm | 40 s | -20 to 50°C | 5 to 95% |
| CO | 0 - 500 | 1 ppm | <± 10% | 1 ppm | 2 s | 0 to 70°C | 5 to 95% |
| H2S (ultra-low) |
0 - 1.0 | TBA | TBA | TBA | TBA | TBA | TBA |
| H2S (low) |
0 - 10 | 10 ppb | <± 0.5 ppm | 0.01 ppm | 2 s | -20 to 50°C | 5 to 95% |
| O3 (ultra-low) |
0 - 0.150 | 1 ppb | <± 0.005 ppm | 0.001 ppm | 65 s | -5 to 50°C | 5 to 95% |
| O3 (low) |
0 - 0.500 | 1 ppb | <± 0.008 ppm < 0.1 ppm <± 10% 0.1 - 0.5ppm |
0.001 ppm | 54 s | -5 to 50°C | 5 to 95% |
| NO2 | 0 - 2 | 0.005 ppm | <± 0.01 ppm up to 0.2 ppm <± 10% > 0.2 ppm |
0.001 ppm | 50 s | -20 to 50°C | 20 to 85% |
| SO2 | 0 - 10 | 0.2 ppm | <± 0.5 ppm | 0.01 ppm | 25 s | -20 to 50°C | 5 to 95% |
| VOC (toluene) |
0 - 500 | 1 ppm | <± 10 ppm < 200 ppm <± 10% 200 - 500 ppm |
1 ppm | 2 s | -20 to 50°C | 5 to 95% |
| CO2 | 0 - 2000 | - | <± 3% | 10 ppm | - | 0 to 50°C | 5 to 95% |
| Temperature | -20 to 100°C | 0.01°C | <± 0.3°C | 0.01°C | - | -40 to 120°C | 0 to 100% |