- 8-12 Hz alpha and 4-8 Hz theta.
- 100-400 Hz EMG filter.
- EMG artifact indicator.
|
Audio: The Alpha-Theta Trainer uses an internal speaker to provide quality audio feedback.
The Alpha-Theta Trainer has two audio feedback modes. In one, the pitch increases or decreases proportionally with the EEG signal. In the other, the audio tones turn on or off to indicate threshold crossing. Feedback can be provided continuously, above or below threshold.
Visual: The alpha and theta frequencies each have separate numerical and lightbar displays (autoranging from 0.1 to 99.9 µV).
You can select to display real-time EEG, time period averaged EEG feedback, or the PERCENTAGE OF TIME the EEG activity is in an alpha or theta state. |
The Alpha-Theta Trainer illuminates an LED and sounds a tone to indicate CROSSOVER (theta >alpha) has occurred.
The Last Value Attained (LVA) feature displays the maximum or minimum EEG value attained (as a target for patient motivation) while the lightbar provides real-time EEG feedback. |
Simply by pressing a Statistics button, the Alpha-Theta Trainer provides both summary and intra-session statistics for the mean, max., min., SD, %SD, and the %Time or Threshold. |
| The AT62 EEG is designed to meet/exceed the electrical and safety standards specified by UL 2601, IEC 601 and CSA 871. |
| Visual & Audio – Per Frequency |
| Two Frequencies |
Alpha only, Theta only, Both Alpha + Theta |
| Display Options |
Numerical LED + Lightbar, Lightbar only, Numerical LED only, A/B Ratio |
| Numerical LED Displays & Range |
Three red digits, 0.5" high; 0.1 - 99.9 microvolts |
| Numerical LED Resolution |
0.10 microvolts |
| TPI (Time Period Integration) Settings |
0.5, 1.0, 2.0, 4.0, 6.0, 10.0, 60.0 seconds |
| Numerical LED Read-Outs |
EEG microvolt, Percentage of time within bandpass |
| LVA (Last Value Attained) Settings |
Continuous, Maximum, Minimum |
| Lightbar Range |
15 segments, x0.1 to x2.5-3.0 proportional scale
|
| Tones |
Piano(default for Alpha), Continuous (default for Theta), Oboe, Flute |
| Volume |
Levels 0 to 7, 4 is the default |
| Provided |
Continuously, Above or Below Threshold |
| Home-Use Software Lockout |
Enabled, Disabled |
| Statistics – Per Frequency |
| Statistics Acquired (Summary) |
Min., Max., Mean, SD, %SD |
| Statistics Acquired (Intra-Stats) |
Choice of Min., Max., Mean, SD, %SD |
| Sampling Capacity |
60 minutes of data |
| Intra-Session Data Collection Rate & Data Display Rate |
0.5 minutes, fixed; 2.5 seconds |
| Statistics Sampling Rate |
4 Hz (every 0.25 seconds) of VRMS |
| Electrical – Per Frequency |
| EEG Bandpass & Rolloff |
8-12 Hz Alpha, 4-8 Hz Theta; 40 db per octave |
| EMG Bandpass & Threshold Range |
100-400 Hz; 0.0 to 99.9 microvolts |
| Common Mode Rejection & Differential Input Impedance |
120 db; >1,000,000 megohms |
| Input Noise & Leakage Current |
0.3 microvolts; In compliance with UL 2601 specifications |
| Audio Output |
3.5 mm stereo earphone jack, (mono output) |
| Analog Output |
Linear 0-5 volt DC signal, compatible with most computer interfaces |
| Line Power Input |
120 volt, 60 HZ, 220 volt, 50/60 Hz |
| Battery Rating |
10 to 12 hours use; 16 hour recharge maximum w/ built-in overcharge protection |
| Dimensions & Weight |
2.5" H x 7.7" W x 5.3" D; 1 pound 11 ounces |
- (1) UL isolation adapter
- (4) Installed Ni-Cad rechargeable batteries
- (1) EEG electrode assembly
- (1) Audio feedback heads
- (1) Carrying case
- (1) Owners manual
|