GENEQ > Biotechnologie > Autoclaves
| Autoclave de laboratoire modèle SN200/300/500 et SQ500 de Yamato |

Model SN200 - 300
and 500 |
Model SQ500 |
| 1.Featured by high performance and easy operation all models are equipped with a coolong fan as standard. |
| • Applicable to protein reformaing at the highest
temperature of 135ºC. • By simple setting and operation, the autoclave manages ordinary sterilization, sterilization of culture mediums and liquids, and melting of cluture mediums. • In addition to default sterilization programs, custom-programmed ones can be saved for repeated use. • To minimize cooling time, a cooling fan is provivded for all models as standard equipment. • The autoclave's operation panel and display are provided on the front part of its lid for easy reading. • With a voltage-sensing power supply mounted as standard equipment,, the autoclave can be connected to any power supplies of 100-120V AC and 200-240 V AC |
| Instrument sterilization program | For sterilization of laboratory instruments, such as flasks, beakers, test tubes and forceps |
| Sterilization and keep warm program | For sterilization and keep warm of culture mediums and liquids, including reagents |
| Melting and keep warm program | For melting and keep warm of solidified agar culture mediums |
| Liquid sterilization program | For sterilization of liquids, such as purified water and dilution water |
| Customer-programmed program | Temperature and processing time can be preset depending user's requirements |
![]() |
|
| 2.Cooling fan as standard equipment |
| To minimize cooling time, a cooling fan is provided for all models as standard equipment. |
| 3.Prepared for GLP/GMP inspections |
| • Two sensor ports are provided as standard on the
chamber. • Recorders and high-performance pressure gauges are available as optional accessories. |
| 4.An optional sample temperature sensor can start sterilization automatically. |
| With the optional mounted sample temperature sensor, you can precisely maintain a desired temperature for the samples to ensure thorough sterilization |
| 5.Front loading drain bottle |
| The drain bottle holder is provided behind the front panel for easy to check the drain water level. |
| 6.Basic safety measures |
| • Steam cup • Lid's interlock mechanism • Self-diagnosis • Alarm against the absence of a drain bottle • Leakage breaker • During any malfunction, the autoclave will automatically shutdown with the alarm buzzer sounding and an error message being displayed. |
| 7.Easy to discharge sterilizing water |
| A large-diameter drainpipe is provided for easier maintenance. |
| 8.Voltage-sensing power unit |
| With a voltage-sensing power unit mounted as standard equipment, the autoclave can be chosen from any power supplies of 100-120V AC or 200-240V AC(single phase). You no longer have to worry about overseas voltage settings. |
| 9.Automatic saving of error history |
| Up to 20 errors can be saved to memory. |
| Specifications | |||||
| Product
code |
241005/241054 |
241006/241055 |
241007/241056 |
241008/241057 |
|
| Model*1
|
SN200/SN210 |
SN300/SN310 |
SN500/SN510 |
SQ500/SQ510
|
|
| System | Automatic high-pressure steam sterilization system for laboratory use *2 | ||||
| Performance | Range of working temperatures | 45-135°C: 45-60°C for preheating, 40-60°C for keep worm, 65-100°C for melting, 105-135°C for sterilization | |||
| Max. working pressure | 0.255MPa | ||||
| Ambient temperature | 5 - 35°C | ||||
| Lid | Manual upward opening with an interlock for safety | ||||
| Main unit | Heater | 100V, 1.2kW (600W 2 units) | 100V, 1.6kW (800W 2 units) | 100V, 1.9kW (950W 2 units) | 100V, 2kW (1000W 2 units) |
| Exhaust valve | An exhaust valve and a slow release valve | ||||
| Connection ports for optional accessories | 1/4" female thread for, sample temp. sensor, chamber temp. sensor, and pressure sensor (branching from the solenoid valve tubing) Total: 3 ports | ||||
| Cooling fan | Axial fan motor | ||||
| Controller | Temperature setting, control and display | PID control by microprocessor, digital setting (by UP/DOWN keys) and display | |||
| Timer and its resolution | 0 or 1 minute to 99 hours 59 minutes in increments/decrements of 1 minute | ||||
| Operation mode | Instrument sterilization course, liquid sterilization course, sterilization and keep/warm course, melting and keep/warm course, customer-programmed course | ||||
| Other features | Key locking, presetting, saving, preheating, forced cooling, sample temperature sensor (option), memory locking, saving the history of up to 20 errors, indication of accumulated working hours, indication of present time, on-off beeping setting | ||||
| Safety measures | Sensor malfunctioning, SSR short-circuit, broken heater wire, prevention of idle heating (liquid expansion type), alarm against the absence of a drain bottle, failure in locking the cover, memory malfunctioning, pressure relief valve (0.255 MPa) | ||||
| Pressure vessel standard | Small-sized pressure vessel (notification of installation is not required) | ||||
| Standard | External dimensions(mm) *3 | 400(W)* 530(D)* 848(H) | 460(W)* 590(D)* 848(H) | 460(W)* 590(D)* 1,058(H) | 520(W)* 660(D)* 876(H) |
| Effective dimensions of chamber (mm) | 240(inner diameter)*445(depth) | 300(inner diameter)*445(depth) | 300(inner diameter)*665(depth) | 370(inner diameter)*410(depth) | |
| Effective Capacity of chamber | 20L | 32L | 47L | 50L | |
| Weight | Approx. 65 kg | Approx. 75 kg | Approx. 85 kg | Approx. 95 kg | |
| Power supply | SN200: 100V-120V AC 12.5A-15A |
SN300: 100V-120V AC 16.5A-20A |
SN500: 100V-120V AC 19.5A-23.5A |
SQ500: 100V-120V AC 20.5A-24.5A |
|
| SN210: 200V-240V AC 6.5A-7.5A |
SN310: 200V-240V AC 8.5A-10A |
SN510: 200V-240V AC 10A-12A |
SQ510: 200V-240V AC 10.5A-12.5A |
||
| Power cord | 3m outside the equipment | ||||
| Accessories | Rack x 2 OMS-60 (dia205xdepth204mm) |
Rack x 2 OMS-70 (dia262xdepth204mm) |
Rack x 3 OMS-70 (dia262xdepth204mm) |
Rack x 2 OMS-90 (dia332xdepth195.5mm) |
|
| Vapor cup, Drain bottle, Drainboard, Filter, Instruction manual | |||||
| Power plug is not provided. *1 Abbreviation (J) is Operation panel board in Japanese and (E) is in English. *2 The autoclaves are not for medical use. *3 External dimensions exclude protrusions. |
|||||
| Optional accessories | ||||
| In addition
to absorbent gauzes, laboratory garments and culturing instruments, a
wide selection of basket and containers for laboratory waste are available. |
![]() |
![]() |
||
| Stainless bucket | ||||
| Product code | Description | Model | Corresponding models | |
| 241087 | Basket (mesh type) | OSM-60 | For SM/SN/SP200 | |
| 241088 | OSM-70 | For SM/SN300/500/SP300 | *1 | |
| 241089 | OSM-80 | For SM/SN/SP500 | *1 | |
| 241099 | OSQ-90 | For SQ500 | ||
| 241093 | Basket (mesh type) with stacking fitting | OSQ-30 | For SM/SN/SP200, with two fittings/set | *1 |
| 241092 | OSQ-40 | For SM/SN/SP300, with two fittings/set | *1 | |
| 241091 | OSQ-50 | For SM/SN/SP500, with three fittings/set | *1 | |
| 241090 | OSR-40 | For SQ500, with two fittings/set | ||
| 241096 | Basket (mesh type) with a mesh shelf | OSQ-60 | For SM/SN/SP200, with 1 mesh shelf | *1 |
| 241095 | OSQ-70 | For SM/SN/SP300, with 1 mesh shelf | *1 | |
| 241094 | OSQ-80 | For SM/SN/SP500, with 2 mesh shelves | *1 | |
| 241097 | OSR-50 | For SQ500, with 1 mesh shelf | ||
| 241083 | Stainless solid basket | OSR-10 | For SM/SN/SP200 | *1 |
| 241084 | OSR-20 | For SM/SN/SP300/500 | *1 | |
| 241098 | OSR-60 | For SQ500 | ||
| 241150 | Stainless bucket | OSN10 | For SM/SN/SP200 | |
| 241151 | OSN12 | For SM/SN/SP300/500 | ||
| 241152 | OSN14 | For SQ500 | ||
| 241153 | Chamber temperature sensor | OSR-70 | For SN/SQ, Type T thermocouple, 3 pieces/set | |
| 241154 | Sample temperature sensor | OSN16 | For SN/SQ, Type T thermocouple, 1 piece | |
| 241155 | Pressure gauge | OSN18 | For SN/SQ, external installation | |
| 241156 | Recorder | RM10C | 6-point type, temp. and pressure recording external installation | |
| 241157 | External alarm output terminal | OSN20 | For SN/SQ | |
| 241158 | Temperature output terminal | OSN22 | For SN/SQ, 4 - 20 mA | |
| 241159 | Time-up output terminal | OSN24 | For SN/SQ, relay, contact output | |