| SPECIFICATIONS |
| Signal source |
monolithic, laser trimmed bridge |
| Operating pressure |
-50 to +300 mmHg |
| Overpressure |
-500 to +4000 mmHg |
| Sensitivity |
5 uV / Volt excitation / mmHg |
| Accuracy |
< ±1.5% or ±1 mmHg (whichever is greater)
|
| Max. excitation |
10 VDC or AC (RMS, < 5 KHz.) |
| Zero offset |
±30 mmHg Max. |
| Zero offset Temp. coefficient |
± 0.25 mmHg / ° C |
| Operating Temp. |
10 to 40° C |
| Volume displacement |
< 0.04 mmHg / 100 mmHg |
| Input impedance |
>500 Ohms |
| Output impedance |
< 900 Ohms |
| Leakage current |
< 10 uA at 115 VAC, 50 Hz. |
| Shock withstand |
3 falls, 1.5 meters onto concrete |
| Max. Storage Temp. |
-25 to +70°C |