GENEQ > Biotechnology > pH Meters

pH/mV/°C Bench Meter
Model pH 300

pH 300 is the basic model of the 300 series and it is manufactured according to CSA, UL and CE standards. It features simplicity of use together with state-of-the-art engineering to measure pH, mV, ORP and temperature. Graphic symbols and messages appear on the LCD to guide the user step-by-step through the calibration procedure, which can be performed at 1 or 2 points from among 3 memorized buffers (pH 4.01, 7.01 and 10.01). A "CFM" message blinks on the LCD until stability is reached, and only then calibration can be performed. This guarantees that only optimal calibrations are carried out.

The calibration date (month, day and year) and data (offset and slope values) are stored into nonvolatile memory, and can be easily recalled by the operator. Erroneous and inaccurate measurements due to non-calibrated system can be avoided with the pH 300 last calibration data recall function, which warns the user in the case of an overdue calibration. With these features pH 300 meets the requirements of Good Laboratory Practice (GLP). Your work can be transferred into a computer via the opto-isolated RS 232 communication port and with the aid of HI 92000, the user friendly Windows® compatible software.

SPECIFICATIONS
Range (pH) 0.00 to 14.00
Range (mV) ±399.9 (ISE); ±1999 (ORP)
Range (°C) -9.9 to 120.0
Resolution (pH) 0.01
Resolution (mV) 0.1 (ISE); 1 (ORP)
Resolution (°C) 0.1
Accuracy (pH)
(@20°C/68°F)
±0.01
Accuracy (mV)
(@20°C/68°F)
±0.2 (ISE); ±1 (ORP)
Accuracy (°C)
(@20°C/68°F)
±0.5
Typical EMC Deviation (pH) ±0.01
Typical EMC Deviation (mV) ±0.6 (ISE); ±1 (ORP)
Typical EMC Deviation (°C) ±0.5
pH Calibration Automatic 1 or 2 points with 3 memorized buffers (pH 4.01, 7.01, 10.01)
Last Calibration Recall Yes
Temperature Compensation Automatic or manual from -9.9 to 120°C (14 to 248°F)
Electrode HI 1131B combination, glass body, refillable, BNC connector (included)
Temperature Probe HI 7669/2W, thermistor type with 1 m (3.3’) cable (included)
Input Impedance 1012 Ohm
Inputs BNC / Reference / ATC
Output Bi-directional RS232 opto-isolated
Power Supply 12VDC adapter (included)
Environment 0 to 50°C (32 to 122°F); RH 95%
Dimensions 280 x 200 x 75 mm (11 x 7.9 x 3”)
Weight 1.3 Kg (2.9 lb.)

CAT. NO. DESCRIPTION
pH 300 Supplied complete with glass body combination HI 1131B pH electrode, HI 7669/2W temperature probe, 12VDC adapter, pH 4.01, 7.01 and 10.01 buffer solutions (20 mL each) and instruction manual.
HI 981000 Comprises cantilever HI 76405 electrode holder, HI 92000 Windows® compatible software, HI 920010/9 PC connection cable, rinsing (5 x 20 mL sachets), storage, cleaning and electrolyte solutions and a refilling pipette.
HI 1131B Combination, refillable pH electrode
HI 3131B Combination platinum ORP electrode
HI 7669/2W Temperature probe
HI 76405 Electrode holder
HI 7004L pH 4.01 buffer solution, 500 mL
HI 7007L pH 7.01 buffer solution, 500 mL
HI 7010L pH 10.01 buffer solution, 500 mL
HI 77400P Cal. kit (pH 4 & 7, 20 mL, 5 pcs each)
HI 710005 110VAC to 12VDC adapter
HI 710006 220VAC to 12VDC adapter
HI 92000 Windows® compatible software
HI 920010 9-pin cable for PC