GENEQ > Biotechnology > Chloride Analyzers
| Chloride Analyzer |
| Model 926S |
Determination of chloride in sweat for diagnosis of fibrosis Cystic. |
| PRINCIPLES OF METHOD |
The main difficulties with the classic method were that silver nitrate reagent would change on storage due to a photochemical reaction and that the titration end point is visually weak. The reagent would have to be standardized requiring a highly skilled analytical technician The 926 method is based on a coulometric titration and is an absolute method where the reagent, the silver ions, is precisely and quantitatively generated at the time of the analysis by passing a constant current between donor electrodes. The end point is detected by the use of sensing electrodes which measure the change in solution conductivity which occurs when excess silver ions are present in solution. |
| REAGENT SYSTEM |
The Buffer system is optimised
to eliminate interfering ions and incorporates a special dispersant to
ensure reproducible formation of the silver chloride precipitate and enhance
analytical performance. |
| SPECIFICATIONS | ||
|
Model 926 |
Model 926S |
|
| Sample size |
500 microlitres |
20 microlitres |
| Readout Range |
10-999mg Cl/l; 2-165mg%Salt |
10-299mmolCl/l |
| Analysis Time |
35 sec @ 200mgl/l |
35 sec@ 100mmolCl/l |
| Accuracy |
SD <1% @ 0.1% Cl Std Solution |
+-2% @ NQCS 100mmol/l |
| Linearity |
1% in range 10-999mgCl/l |
1% 10-299 mmolCl/l |
| Reproducibility CV for 20 replicates |
<1.5% @ 200mgCl/l |
<1.5% @ 100mmolCl/l |
| REAGENT SYSTEM |
The Buffer system is optimised to eliminate interfering ions and incorporates a special dispersant to ensure reproducible formation of the silver chloride precipitate and enhance analytical performance. |
| THE SHERWOOD ADVANTAGE |
|