GENEQ > Biotechnology > Pumps

Compulab® 3 Dilutor/Pump System
 


• Programmable to automate your fluid handling processes
• Operate up to four pumps independently from one keypad or your PC

Product Description
Standard Fluid Transfer:
• One bench top programmable controller capable of running up to four individual RD1 or RD2 pumps synchronized and independently. (Mix-or-match)
• Flow rate of RD-1 pump is .4-3450 ml/min. The RD-2 pumps .5-5000 ml/min.
• Easy connection of control and pump drives with supplied 3 1/2 ft RJ-12 (phone) cord. Maximum connection length of 15 ft.
• Up to 30 user programs can be stored in memory.
• Accuracy and Repeatability ±1%.
• 1 to 999 cycles or continuous operation in dispense and dilute modes.
• Manual or automatic timed delay cycling.
• Interval delay time ranges from as low as 0.1 seconds to 2.780 hours programmed in 0.1 second increments.
• User-friendly, menu-driven keypad with 80 character, backlit, alphanumeric, liquid-crystal display (2 x 40).
• Twenty-key keypad facilitates direct data entry. Keypad provides tactile feedback.
• No calibration necessary when changing volume with the same size tubing.
• Both pump models are dual voltage 115/230 VAC (switchable) and meet UL 508, IEC 1010, cUL and CE regulations. The controller power is supplied by the pump drives.
• Speed Control Feedback maintains accurate flow rate. 1/10 HP motor provides controlled responsive dispensing.
• Direct input of volumes, flow rates, and speed. Ranges determined by tubing size.
• Pump speed and dispense volume can be changed "on-the-fly" while pump is running.
• Direct connection and control from a PC- compatible DB9 serial port.
• The pump drives have an illuminated On/Off switch and
seven-segment LED to indicate the pump address.
• RD1 and RD2 drives come complete with chemically resistant pump head (molded of glass filled Noryl®). The RD1 pump head has three stainless steel rollers. The RD2 pump head has three Delrin rollers.
• Operating temperature 0°C to 40°C.
• IP rating: 22
Dispensing/Diluting:
• Dispense delivery is set by volume, not limited to time, from 0.5 mL to 5000/L for RD-1 and 0.5 mL to 7000/L for RD-2 based on tubing size.
• Programmable anti-drip feature reverses pump for set rotations after each cycle. Ideal for more viscous solutions.
• Completed dispense cycles are confirmed with an audible prompt programmable for 0.0 to 60 seconds. Semi-automatic dispenses are individually prompted.
• Direct connection and control in conjunction with Ohaus and Sartorius balances.

Connect 4 Pump Drives
to a Single Compulab® Controlle
r
Check the charts below for the pump drive(s) that best suit your application. ùorder the controller and any combination of pump drives (up to four)


Flow Rates in ml / min
Cat. No
Description
Tubing
Size 13
Size 14
Size 16
Size 25
Size 17
Size 18
72-666-100
RD2 Pump
ml / min
.5-50
3-300
12-1200
26-2600
42-4200
50-5000
72-777-100
RDI Pump
ml / min
.4-40
2-200
7.5-750
17-1700
26.5-2650
34.5-3450

Dispensing volumes ml / cycle
Cat. No
Description
Tubing
Size 13
Size 14
Size 16
Size 25
Size 17
Size 18
72-666-100
RD2 Pump
ml / min
.5-50
3-300
12-1200
26-2600
42-4200
50-5000
72-777-100
RDI Pump
ml / min
.4-40
2-200
7.5-750
17-1700
26.5-2650
34.5-3450

Specifications and ordering information
Compulab®3 Controller (includes software)
Cat. No
Dimensions (LxWxH)
Shipping Wt.
Power
72-767-100
8-3/4" x 6-1/8" x 3-1/4"
front to 7-3/4" back
5 lbs
Powered by Pump Drive
2.7 kg

Compulab®3 Pump Drives (includes dispensing wand)
Cat. No
Pump
Drives
RPM
Dimensions
Speed
Motor
IP Rating
LxWxH
Shipping
Power
Control
Size
 
(Pump Drive)
Weight
50/60 Hz
72-666-100
RD2 Pump
ml / min
.5-50
3-300
12-1200
26-2600
42-4200
50-5000
72-777-100
RDI Pump
ml / min
.4-40
2-200
7.5-750
17-1700
26.5-2650
34.5-3450

Tubing Cross Sections
Size 13
Size 14
Size 16
Size 25
Size 17
Size 18