GENEQ > Environnement > Conductimètres
| TDSTestrs™ et ECTestrs™ |
| Microprocessor-based, imperméable |
|
![]() |
|
| These waterproof and
dustproof testers offer ±1% full scale accuracy and push-button calibration! |
||
| Waterproof, dustproof housing Meets IP67 ratings for highest possible dustproof and waterproof protection. Plus, there’s no need to worry if you drop your tester into water—it floats! Replaceable electrode you can use the same meter body over and over again, which saves you money Push-button calibration calibrate your tester more precisely than trimpot adjustment; no screwdrivers necessary HOLD function freezes reading until you can log it Auto shut-off turns off tester after 8.5 minutes of nonuse; saves your batteries Four models available TDSTestr low: 0 to 1990 ppm (10 ppm resolution) TDSTestr high: 0 to 10.00 ppt (0.10 ppt resolution) ECTestr low: 0 to 1990 µS (10 µS resolution) ECTestr high: 0 to 19.90 mS (0.10 mS resolution) Full reading displayed no need to multiply the readout to obtain actual test values Automatic temperature compensation (ATC) gives you accurate readings even in fluctuating temperatures Calibration kits available Featuring everything you need for calibration, measurement and storage of your waterproof TDSTestr |
||
| Clockwise from near
right: TDSTestr low shown actual size • Calibration kit includes everything you need for fast, accurate calibration • Calibration button tucked away in battery compartment to avoid accidental recalibration • Replaceable electrodes extend tester life • Testr carrying case stores, transports and protects your tester in the field • Choose from four models for conductivity or TDS measurements |
||
| APPLICATIONS |
||
| • TDSTestr low: Use
for measuring TDS in natural water. Verify reverse osmosis system operation
and tap water quality. Check nutrient solution concentration in hydroponics applications. • TDSTestr high: Use for measuring TDS in salt water, waste water, cooling tower water, and boiler condensate. Check nutrient solution concentration in hydroponics applications. Measure salinity for ponds and recirculating systems. Check saline and chemical levels in pools and spas. • ECTestr low: Use for measuring conductivity (µS) in natural water. Verify reverse osmosis system operation and tap water quality. Check nutrient solution concentration in hydroponics applications. • ECTestr high: Use for measuring conductivity (µS) in salt water, waste water, cooling tower water, and boiler condensate. Check nutrient solution concentration in hydroponics applications. Measure salinity for ponds and recirculating systems. Check saline and chemical levels in pools and spas. |
||
| SPECIFICATIONS | ||||
| Model | waterproof
TDSTestr low |
waterproof
TDSTestr high |
waterproof
ECTestr low |
waterproof
ECTestr high |
| Range | 0 to 1990 ppm |
0 to 10.00 ppt |
0 to 1990 µS |
0 to 19.90 mS |
| Resolution | 10 ppm |
0.10 ppt |
10 µS |
0.10 mS |
| Accuracy | ±1%
full-scale |
|||
| Calibration | One point
push-button calibration (using keys inside battery compartment) |
|||
| Calibration standard range |
300 to 1990 ppm |
3 to 10.00 ppt |
300 to 1990 µS |
3 to 19.90 mS |
| Operating temp. | 0 to
50°C; 32 to 122°F |
|||
| Temperature compensation |
Automatic
(ATC), 0 to 50°C, 32 to 122°F |
|||
| ATC coefficient | Beta=2%
per °C, 1.11% per °F, 25°C reference temperature |
|||
| Power and battery life |
Four
1.5 V alkaline batteries Eveready A76BP (supplied) 100 hrs. continuous
use (approx. 600 tests per battery pack). Alternate replacement model Eveready 357BP silver oxide, 140 hrs. continuous use. |
|||
| Dimensions | Unit
only: 61⁄2"L x 11⁄2" dia. (165 x 38 mm); Boxed:
71⁄4" x 23⁄4" x 17⁄8" (184 x 70 x 48
mm) |
|||
| Weight | Unit
only: 3.25 oz (90 g); Boxed: 6.0 oz (170 g) |
|||
| Wetted materials | 316 Stainless
Steel (electrodes) and glass filled thermoplastic polyester |
|||
| ORDERING INFORMATION | |
| All TDSTestrs include batteries and operating instructions. | |
| WD-35661-11 Waterproof TDSTestr low 0 to 1990 ppm |
|
| WD-35661-41 Waterproof ECTestr high 0 to 19.90 mS |
|
| ACCESSORIES | |
| WD-35661-17 Replacement electrode sensor,
for all TDSTestrs and ECTestrs listed above. WD-35624-45 Waterproof tester belt loop vinyl carrying case—holds one waterproof tester. Features separate pocket to hold calibration solution pouches. Shpg wt 0.5 lbs (227 g) WD-09377-16 Replacement batteries, 1.5 V, 6/pack WD-35661-70 Deluxe TDSTestr kit. Includes 2 each calibration pouches 447 µS, 1413 µS, 2764 µS, 15,000 µS, and rinse water; sample jar; miniature screwdriver; foam-lined plastic carrying case; and instructions. Order TDSTestr separately. Shpg wt 0.7 lb (300 g) |
|
| CALIBRATIONS SOLUTIONS | ||||
| Conductivity-to-TDS conversion
factors |
||||
| Catalog Number |
Calibration solution type |
ppm KCl |
ppm NaCl |
ppm 442 |
Calibration
solution pouches. box of 20. Each pouch contains 20 ml of premixed calibration
solution. Since the pouches are individually sealed, the solution is guaranteed to be fresh and contaminant free every time. Accuracy ±1% at 25°C. Shpg wt 1 lb. (454 g). |
||||
WD-35653-10 |
447 µS |
225.6 |
215.5 |
300.0 |
WD-35653-11 |
1413 µS |
744.7 |
702.1 |
1000 |
WD-35653-12 |
2764 µS |
1382 |
1414.8 |
2062.7 |
WD-35653-13 |
15,000 µS |
8759 |
8532 |
13,455 |
| Calibration solution bottles. 1 pint of premixed calibration solution in recyclable-PE bottle. Accuracy ±1% at 25°C. Shpg wt 1.1 lbs (510 g). | ||||
WD-00653-16 |
84 µS |
40.38 |
38.04 |
50.50 |
WD-00653-47 |
447 µS |
225.6 |
215.5 |
300.0 |
WD-00653-18 |
1413 µS |
744.7 |
702.1 |
1000 |
WD-00653-15 |
1500 µS |
757.1 |
737.1 |
1050 |
WD-00653-27 |
2070 µS |
1045 |
1041 |
1500 |
WD-00653-20 |
2764 µS |
1382 |
1414.8 |
2062.7 |
WD-00653-89 |
8974 µS |
5101 |
4487 |
7608 |
WD-00606-10 |
12,880 µS |
7447 |
7230 |
11,367 |
WD-00653-50 |
15,000 µS |
8759 |
8532 |
13,455 |
| For our standard waterproof TDSTestrs, request bulletin CON A3. For our + Series waterproof TDSTestrs and ECTestrs, request bulletin CON A5. | ||||