GENEQ > Environment > Dissolved Oxygen Meters
Dissolved Oxygen Meter
Model Acorn DO6
![]() |
| FEATURES |
|
![]() |
![]() |
![]() |
![]() |
|
Large customized LCD
|
Splash-proof keypad
|
Battery compartment
|
Rubber boot with tilt stand
|
| % SATURATION OR CONCENTRATION MODE |
| At the push of a button, the meter switches measurement mode from mg/l or ppm to % saturation. |
| USER CUSTOMIZATION |
| Custornization functions include: % saturation, mg/I or ppm mode, hold, auto-off, barometric and salinity correction, automatic temperature compensation, electrode diagnosis, user reset and restore to factory defaults. |
| PRESSURE AND SALINITY COMPENSATION |
| Barometric pressure and salinity values can be manually input in the meter. The meter automatically adjusts for necessary compensation, thus eliminating the need for cumbersome charts. |
| EASE IN CALIBRATION |
| For 100% or 0% saturation calibration, simply press the CAL button. Calibration for mg/I or ppm modes can also be performed using a known oxygen concentration solution or reference DO meter. |
| GALVANIC ELECTRODE WITH BUILT-IN TEMPERATURE SENSOR |
| Galvanic cell technology allows instantaneous measurements without waiting for the probe to be polarized as in the case with amperometric electrodes. The rugged galvanic electrode is optimized for stable response and is able to withstand the rigors of portable use. It requires little or no maintenance and has a built-in sensor for automatic temperature compensation. Membrane replacement is simplified by our single-piece membrane cap. |
| ELECTRODE DIAGNOSIS |
| Condition of electrode can be viewed through the setup menu. This makes it easier for the user to check the efficiency of the probe. |
| INNOVATIVE DESIGN |
| The Acorn' DO 6 meter is ergonomically designed for convenient one-handed operation. The ribbed body ensures firm grip preventing meter from slippage. |
| PROTECTIVE RUBBER BOOT |
| A sturdy rubber boot protects the meter from damage. The tilt stand at the back of the boot allows for easy benchtop viewing. |
| RUGGED, ALL-WEATHER CARRYING CASE |
| A custom-built rugged carrying case provides convenient storage of meter, electrode, refilling solutions and accessorieshelpful for field testing. |
| WARRANTY |
| Free from defects in materials and workmanship for I year and its electrodes to be free from defects in materials and workmanship for 6 months. See product packaging for full warranty terms. |
| SPECIFICATIONS | |
| % Saturation of Oxygen | |
| Range | 0.0 to 200.0% |
| Resolution | 0.1% |
| Accuracy | ±1.5% of full scale |
| Dissolved Oxygen | |
| Range | 0.00 to 20.00 mg/l or ppm |
| Resolution | 0.01 mg/l or ppm |
| Accuracy | ±1.5% of full scale |
| Temperature | |
| Range | -5.0 to 105.01 C |
| Resolution | O.I*C |
| Accuracy | :tO.5'C |
| Barometric Pressure | |
| Range | 500 to 1499 nun Hg |
| Correction | 66.6 to 199.9 kPa |
| Resolution | 1 mm Hg or 0.1 kPa |
| Method | Automatic correction after manual input |
| Salinity Correction | |
| Range | 0.0 to 50.0 ppt |
| Resolution | 0.1 ppt |
| Method | Automatic correction after manual input |
| General | |
| Temperature Compensation | Automatic / Manual from 0 to 50*C |
| Operating Temperature | 0 to 50*C |
| Probe | Galvanic |
| Response Time | 60 seconds to achieve 95% of reading |
| HOLD function | Yes |
| Auto-off function | Selectable (20 minutes after last key press) |
| Display | Custom single 4-digit LCD |
| Inputs | BNC for DO and 2.5 mm phono for temperature |
| Power Requirements | Four AAA Batteries included |
| Battery Life | >700 hours (alkaline batteries) |
| Dimensions (meter) | 14 (L) x 7 (W) x 3.5 (H) cm |
| Weight (meter) | 200 g |
| Dimensions (probe) | 115 (L) x 12 (dia) mm with 1-meter cable |
| Dimensions (case) | 27 x 35 x 7.5 cm |
| Weight (case) | 1.7 kg |
| CAT. NO. | DESCRIPTION |
| OAWD-35642-10 | Acorn DO 6 dissolved oxygen portable meter with glavanic DO probe and carrying case |
| OAWD-35642-50 | Galvanic dissolved oxygen probe for Acorn DO 6 meter (115 x 12mm , LxDia.) with 3-ft cable, 10-ml bottle of electrolyte solution and spare pre-membrane housing cap |
| OAWD-35640-71 | DO probe electrolyte solution, 50ml |
| OAWD-35642-55 | Maintenance kit; includes replacement cap and 10ml of electrolyte solution |