| Enregistreur d'événements et compteur |
| Modèle EL-USB-5 |
![]() |
![]() |
| Features |
• Functions as an event logger, state change logger and event counter • Logging rate of up to four times per second for events, two times per second for state changes and 100 times per second when counting • Stores up to 32,510 readings in event mode • Stores up to 32,510 readings in state change mode • Stores up to 32,510 timed intervals in event counter mode • Accepts input voltages of up to 24V • Volt free contacts option (normally open contact only) • Rising and falling edge triggering • LED flash on event / state change • Connection via two screw terminals • USB Interface for set-up and data download • Supplied with replaceable internal lithium battery, Windows control software and measurement lead |
This standalone data logger can record events (detecting an input and storing the time and date), record state changes (similar to an ‘event’ but also recording when the input reverts back to normal) and count events (count the number of events that happen during a timed interval). |
| Specifications |
| Specifications | Minimum | Typical | Maximum | Unit |
| Time between events | 200 | Milliseconds | ||
| Time between state changes | 500 | Milliseconds | ||
| Time between event counts | 10* | Milliseconds | ||
| Input voltage ** | 3 | 28 | V d.c. | |
| 'Volt free contacts' voltage *** | 2.75 | 3.6 | V | |
| 'Volt free contacts' current | 35 | μA | ||
| Timing accuracy† | ±3 | Seconds per 24 hours | ||
| Operating temperature range | -35 (-31) | +80(176) | °C (°F) | |
| 1/2AA 3.6V Lithium Battery Life‡ | Years | |||
* 10 milliseconds can only be achieved with the LEDs disabled. If the LEDs are enabled, the time between the event counts is reduced to 50 milliseconds |
||||
| EL-WIN-USB (Control Software) |
Lascar’s EasyLog USB control software is supplied free of charge with each data logger. Easy to install and use, the control software runs under Windows 2000, XP, Vista & 7. The software is used to set-up the data logger as well as download, graph and export data to Excel. |
||||
The software allows the following parameters to be configured: • Logger name • Record events, record state changes or count events • Triggering on rising of falling edges (see ‘Edge Triggering’ section) • Voltage or Volt free contact triggering • Flash LED on event/state change • Voltage range • Time period for event counting (i.e count the number of events every 10s). Maximum number of events per time period is limited to 65,000. |
![]() |
|||
| LED Flashing Modes |
![]() |
Green single flash (10 seconds) |
Green single flash (20 seconds) |
|
Green double flash (20 seconds) |
|
Red single flash |
| Voltage Input Range |
The logger features a selectable voltage input range, which allows the logger to operate with a wide variety of system. Each input range has a different high and low ‘trigger voltage’ (i.e the voltage at which the logger considers a signal to be high or low). |
||
| Voltage range | Low trigger volatge | High trigger voltage |
| 0 - 3V | 0.8V | 2V |
| 0- 5V | 1.3V | 3.4V |
| 0 - 12V | 3.2V | 8V |
| 0 - 24V | 6.4V | 16V |
| Edge Triggering |
If the user selects ‘rising edge’ triggering, then an event is recorded at the point a signal goes from low to high. This is the default setting. |
![]() |
| If the user selects ‘falling edge’ triggering, then an event is recorded at the point a signal goes from high to low. |
![]() |
| Battery Replacement |
We recommend that you replace the battery every 12 months, or prior to logging critical data. WARNING |
| Dimensions |
All dimensions in mm (inches) |
| The Easylog USB Range |
Each EL-USB data logger features the direct-to-USB connection and easy-to-use functionality that the range is known for. The range comprises of the data loggers as detailed in the following table: |
|||||||
| Part No. | Function | Range | Accuracy (overall error) | Readings | Battery | Battery Life* | |
| Typ. | Max. | ||||||
| EL-USB-1 | Temperature | -35 to +80°C (-31 to +176°F | ±1ºC (±2ºF) | 16,382 | 3.6V ½AA | 1 Year | |
| EL-USB-1-PRO | High Temperature | -40 to +125°C (-40 to +257°F | ±0.2ºC (±0.4ºF) | ±0.5ºC (±1ºF) | 32,510 | 3.6V ?AA | 3 years |
| EL-USB-1-RCG | Temperature with rechargeable battery | -20 to +60°C (-4 to +140°F | ±1ºC (±2ºF) | 32,510 | Lithium Ion | 1 months (rechargeable) | |
| EL-USB-2 | Temperature, humidityand dew point | -35 to +80°C (-31 to +176°F) |
±0.5ºC (±1ºF) |
±2ºC (±4ºF) |
16,382 | 3.6V ½AA | 1 year |
| EL-USB-2+ | Increased accuracy temperature, humidity & dew point | -35 to +80°C (-31 to +176°F) |
±0.3ºC (±0.6ºF) |
±1.5ºC (±3ºF) |
16,382 | 3.6V ½AA | 1 year |
| EL-USB-2-LCD | Temperature, humidity & dew point with LCD | -35 to +80°C (-31 to +176°F) |
±0.5ºC (±1ºF) |
±2ºC (±4ºF) |
16,379 | 3.6V ½AA | 1 year |
| EL-USB-2-LCD+ | Increased accuracy temperature, humidity & dewpoint with LCD | -35 to +80°C (-31 to +176°F) |
±0.3ºC (±0.6ºF) |
±1.5ºC (±3ºF) |
16,379 | 3.6V ½AA | 1 year |
| EL-USB-3 | Voltage | 0 to 30V d.c. | ±1% | 32,510 | 3.6V ½AA | 1 year | |
| EL-USB-4 | Current loop | 4 to 20mA | ±1% | 32,510 | 3.6V ½AA | 1 year | |
| EL-USB-5 | Counter, Event & State | N/A | ±3 secs/24 hrs | 32,510 | 3.6V ½AA | 1 year | |
| EL-USB-TC | Thermocouple (J, K and T-type) K-type probe included | -200 to +1350ºC (-328 to +2462ºF) (K-type) -200 to +1190ºC (-328 to +2174ºF) (J-type) -200 to +390ºC (-328 to +734ºF) (T-type) |
±1ºC (±2ºF) | 32,510 | 3.6V ½AA | 6 months | |
| EL-USB-TC-LCD | Thermocouple with LCD (J, K and T-type) K-type probe included | -200 to +1350ºC (-328 to +2462ºF) (K-type) |
±1ºC (±2ºF) | 32,510 | 3.6V ½AA | 6 months | |
| EL-USB-CO | Carbon monoxide | 0 to 1000ppm NOT A LIFE SAVING DEVICE | ±6ppm | 32,510 | 3.6V ½AA | 3 months | |
| EL-USB-CO300 | Carbon monoxide | 0 to 300ppm NOT A LIFE SAVING DEVICE | ±4ppm | 32,510 | 3.6V ½AA | 3 months | |
| EL-USB-LITE | Low cost temperature | -10°C to +50°C (+14 to +122°F) | ±1ºC (±2ºF) | 4,080 | CR1620 |
1 months | |
| EL-USB-RT | Real-time temperature & humidity monitor | -20 to +70°C (-4 to +158°F) | ±1.5ºC (±3ºF) |
7 days | N/A | N/A | |