GENEQ > Meteorology > Radiation Shield

Fan Aspirated Radiation Shield
Model 076B

 

Radiation Shields are an important component of accurate air temperature and humidity measurements. The function of the radiation shield is to protect the sensor from direct and indirect radiant energy, which could cause unknown and uncontrolled errors. The Met One Instruments Model 076B is suitable for the most stringent applications.
Features
  • Errors reduced to < 0.05°F
  • Accommodate several sensors for Delta T measurements
  • Radiation errors independent of outside emissivity
  • Low-conductive interconnects
  • DC or AC power operation
  • Gloss white finish
  • Corrosion resistant materials throughout
  • Easy mounting and tower servicing

The Model 076B Fan Aspirated Radiation Shield virtually eliminates errors caused by solar or terrestrial radiation, as well as secondary errors caused by convective heat transfer from the outer shield surfaces. It is designed to continuously sample ambient air for high accuracy temperature, differential temperature, relative humidity and dewpoint measurements. Actual test results showed errors of less than 0.05°F (Test Report available).

The Model 076B is designed for easy maintenance and sensor calibration. The lower section, containing the sensors, can be easily removed from the shield/fan assembly. Individual sensors, connected to the system through a sealed junction box, are easily replaced, serviced, or calibrated from ground level.

Operation
Air is drawn into the bottom of the sensor housing through two concentric ports. High velocity air through the outer port scrubs the outer surfaces and is exhausted. The inner port draws in the air sample, enabling a true gas temperature measurement to be made. This air flow system prevents convective heat transfer to the sampling stream.
Construction
The shield is constructed of aluminum and other corrosion-resistant materials. All outer surfaces have been painted with high-gloss epoxy enamel for maximum reflectivity. All interconnecting materials have been selected to minimize conductive heat transfer. The combination of air flow, symmetrical shield design, and surface finish provides a shield that virtually eliminates all errors caused by radiation.
Specifications
Operating Characteristics
Aspiration:
Electric Fan
Flow Rate:
1000 CFM Scrubbing
500 CFM Sample
Temperature Range:
Standard
-50°C to +85°C
Optional -50°C to +100°C
Power Required: 115 VAC 20 watts
12 VDC 250 mA
220 VAC 18 watts
Physical Characteristics
Material:
Aluminum
Finish:
White Epoxy
Dimensions:
20 in (51 cm) h x 20 in (51 cm) diam.
Weight, less sensor(s): 9.3 lbs (4.2 kg)
Sensors
Maximum Number:
4
Models:
060/062/066/083
Maximum Diameter:
.75 in (1.9 cm)
Maximum Length:
10 in (25 cm)
Transducer Sensing Zone: 8 in (20 cm)
Mounting
Shield mounts to 3/4 in. IPS horizontal pipe
Ordering Information
Model 076B-1: 115 VAC
Model 076B-4:
12 VDC
Model 076B-11: 220 VAC
Order model number corresponding to voltage requirement.
Consult factory for installations requiring dew point measurements utilizing the Model 078 sensor.
Cable:
Signal
PN 2144-xx (xx = length in ft)
Power PN 1954-xx (AC)
PN 2423-xx (DC)