- Large, easy to read LCD display
- Trip history statistics available on LCD
- Single-use
- USB Connection
- NIST traceable and CE Mark
- IP54 water-resistant rating
- Print data using Graphics Printer, FlashPDF software or download to PC
- Patented Shadow Log feature guarantees data even if the operator fails to start the unit
|
| Measurement Range |
-40°F to 149°F (-40°C to 65°C) |
| Temperature Resolution |
0.1°F/°C |
| Accuracy |
±1.0°F from 14°F to 77°F (±.5°C from -10°C to 25°C);
±2.0°F above 77°F (±1.0°C above 25°C);
±2.0°F below 14°F (±1.0°C below -10°C) |
| Data Capacity |
3823 points |
| Data Download Time |
2 seconds typical for 3823 points |
| Factory Programable Alarm Levels |
High & Low levels over entire range, displayed on LCD |
| Battery |
(CR2032) 3.6V Lithium/18 months |
| Dimensions/Body |
1.575" x 3.189" x .575" (40mm x 81mm x 15mm) |
| Casing |
PC/ABS plastic |
| Weight |
1.3 oz. (36 grams) |
| Shock Resistance |
Withstands 3-foot (1 meter) drop to hard surface |
| Certification |
CE Mark, NIST traceable, ROHS Compliant |
| Display |
LCD, Red LED Alarm |
| Waterproof Rating |
IP54 |
| Storage conditions |
-4°F to 122°F (-20°C to 50°C) 0% to 95% RH w/o condensation |
| Min. System Requirements |
Windows PC with XP, Vista or Windows 7 operating system |
| Download Requirements |
Direct USB connection to PC. View data with free software downloaded from our web site. |
| DeltaTRAK manufactures products under an ISO 9000 registered quality management system |