| THE MAIN COMPONENTS |
| Chatillon DFGS series gauges in the following
capacities |
| DFGS-100 |
.100 LB. x .05 LB. FGS-50 KG |
.50 KG x .05 KG |
|
| DFGS-50 |
.50 LB. x .02 LB. FGS-25 KG |
.25 KG x .01 KG |
| Stainless steel base |
| Stainless steel base which includes a vertical column to
which the pneumatic cylinder and gauge are mounted. The base has a filter-regulator
with gauge to which the plant air supply is connected. The base also has
two palm button actuator valves. |
| The plunge |
| The plunger and plunger ring plate are machined in accordance
with ASTM Spec. D4032. |
| Template to cut |
| Template to cut 4" X 8" fabric sample. |
Cut a sample of fabric not less than 4 inches wide
in the direction of fabric warp. Sample swatch must be taken at least 4 inches
from the selvage. From this swatch, use 4" X 8" template to cut three
samples. Each sample must be free of wrinkles and creases. Mark the face side
of each sample. Fold each sample back so the double ply measures 4" X 4".
Place the folded sample on the plunger ring plate
making sure the sample is centered on the plate. Press "ON" to turn
gauge on. Select units by pressing "UNITS" until the desired units
appear on display. Select "PEAK" mode of operation. Press "ZERO"
to zero the display. Make sure sample is centered on the plunger ring plate
and press both palm buttons simultaneously to push sample through the plunger
ring plate. Palm button must be held down until sample is pushed completely
through plunger ring plate. Using two palm buttons insures operator safety and
prevents erroneous readings caused by touching sample with free hand.
Record the gauge reading for each sample and report
the average of the three samples. Discard the results of any sample, which varies
more than 20% from the average of the three samples, and replace it with a new
sample.
|