The 5.0 nm spectral bandwidth is ideal for
most routine quality control measurements.
When high resolution is not required to
resolve closely spaced peaks, the added
energy throughput of the GENESYS 10S Vis
extends the sensitivity of your analysis and
increases the signal-to-noise ratio for dilute
samples. Precision electronics and a simple,
single-beam optical geometry provide
accurate results.
Power and Performance With or Without a PC
The GENESYS 10S series instruments are
optimized for the classroom or teaching lab –
offering great value and many features
for their small footprint. These lightweight
instruments are easy to transport and store.
Offering the ergonomic choice of a simple,
but powerful embedded software interface or
optional computer control with application
software tailors the spectrophotometer
to your needs. Use the embedded software
to acquire scanning, fixed wavelength,
kinetics, or concentration data and easily
save the experi mental data in a spreadsheet-
compatible, CSV format.
Students can:
- acquire the data in the lab
- save it to a memory device
- use a separate computer for analyzing
data and writing lab reports
Application software with computer
control of the instrument further extends the
types of experiments that can be performed in
the laboratory and allows more sophisticated
analysis. VISIONlite instrument control
software offers an intuitive interface for
wavelength scanning, fixed wavelength
analysis, single and multi-cell kinetics, and
quantitative analysis. This easy-to-use and
easy-to-teach interface makes acquiring,
analyzing and exporting data straightforward– allowing you to focus on the experiment,
not the software.
Automation for Routine Analysis
With a small footprint and easy-to-use
embedded software, the GENESYS 10S
series is ideal for routine sample analysis.
For simple, multi-sample experiments,
automate your analysis with the built-in
6-Cell Changer. With one position reserved
for a blank, the cell changer allows the
automated analysis of up to five samples. The
6-Cell Changer can be removed easily and
replaced with other cell holders including
long pathlength cell holders, Peltier and
recirculating water temperature control cell
holders, and flow cells for sipper systems.
For high throughput liquid sample
analysis, a versatile peristaltic pump sipper
system is available. Installation is fast and
simple – slide on the sipper spout, attach
tubing and you are ready to take data.
A variety of flow cells is available, with
pathlengths from 1 to 100 mm and volumes
as low as 8 microliters.
Performance Verification
Performance verification tests included in the
embedded software of every GENESYS 10S
series instrument provide pre-programmed
methods for verifying instrument performance.
In accordance with GLP and GMP, each
verification report gives the time, date, and
instrument serial number. The built-in wavelength
accuracy test is compatible with either
the internal lamp or external calibrated
standards. Additional built-in tests allow
you to monitor instrument performance to
ensure reliable data collection.
The xenon lamp of
the GENESYS 10S UV-Vis
provides an internal
standard for wavelength
accuracy verification. To
validate the instrument
performance further,
built-in test methods for
stray light, noise, and
resolution are available.
A cost-effective Green Dye standard
is available to verify both photometric
accuracy and wavelength accuracy of the
GENESYS 10S series instruments. With
certified values at 260, 430, and 630 nm,
this standard provides a check of wavelength
and photometric accuracy from the
UV to the near-IR. |