- IP 570 designated new Energy Institute (IP) Standard Test Method has been developed
- Measurement range from 0-250 mg/kg H2S (0-250 ppm H2S) in the liquid phase
- Fast measurement time of less than 15 minutes compared to 3 hours with IP 399
- A non chemical method, no wet chemistry involved
- Small lab bench footprint, fully portable operation
- Ideal for monitoring residual fuel blends, cargoes & products in the distribution system
- Critical Measurement method for Product Safety and Release
|
• Refineries and Fuel Blending Locations
• Tank Storage Terminals
• Marine Fuel Suppliers
• Inspection Companies
• Independent Analytical Services
|
.jpg) |
| A small volume of sample is dissolved in a diluent and heated under precisely controlled conditions to
60°C to release any entrained H2S content present in the sample. Clean air is passed through the
sample vessel to carry the H2S to the sensor, where the concentration is measured by an electrochemical
sensor. Airfflow through the sensor is measured by a mass flow meter. When the H2S
concentration in the sample has been fully driven off the Analyser will calculate and report the total
H2S released from the sample. |
| To conduct a test, 20ml of diluent is decanted into the Sample Vessel, which is inserted into the Heater
Block. After approximately 5 minutes, the diluent has reached 60°C and 1ml of sample is added to the
diluent. Sample identity,operator name, weights of empty/charged syringe are input via the menu
display and keys on the front membrane panel. |
| The test is initiated by pressing the START/STOP Button, sample analysis is fully automated and result
results are stored to memory at the end of the test. The Analyser software automatically detects leaks
in the system and produces error codes and alerts if sample/test analysis is void. |
.jpg) |
| Measurement range |
0-250 mg/kg H2S in the liquid phase (0-250 ppm H2S) |
| Operating limits |
5 – 40c maximum (80% RH) |
| Viscosity Range |
Up to 3000 mm2/s |
| Principle of measurement |
Advanced Electrochemical sensor |
| Test duration |
15 minutes |
| Sample size |
1ml, 2ml, 5ml (depending on H2S concentration) |
| Diluent volume |
20 ml |
| Voltage |
12V DC, supplied with universal A/C transformer |
| Power |
60W max |
| Computer interface |
RS232 |
| Size (H x W x D) |
210 x 300 x 410 mm |
| Weight |
8 kg |