Geneq > Environnement > Turbidité - Mesure

Turbidimètre portatif HACH
Modèle 2100Q


High Accuracy Turbidity Measurement

The 2100Q Portable Turbidimeter offers a unique combination of advanced features, such as easy calibration and simplified data transfer, and measurement innovation for rapidity settling samples, giving you confidence you're getting accurate results every time.

Easy On-Screen Assisted Calibration and Verification

Save time and get accurate results with an easy-to-follow interface that eliminates the need for complicated manuals to perform routine calibrations. Single-standard RapidCal™ calibration offers a simplified solution for low level measurements, while ensuring you meet reporting requirements.

Simple Data Transfer

Customizable power and connectivity modules enable smooth data transfer and flexibility. Optional USB + Power module allows data downloads to any computer via USB port.

Accuracy for Rapidity Settling Samples

The innovative Rapidly Settling Turbidity™ mode provides accurate measurements for difficult to measure, rapidly settling samples. An exclusive algorithm that calculates turbidity based on a series of automatic readings eliminates redundant measurements and estimating.

Convenient Data Logging

Up to 500 measurements are automatically stored for easy access and backup, including: date and time, operator ID, reading mode, sample ID, sample number, units, calibration time, calibration status, error messages and the result.


Features
  • On-screen, assisted operation
  • Single Step Calibration Function
  • Rapidcal™ Single Standard Calibration
  • RST Feature for improved accuracy and precision with rapidly settling samples
  • Data log
  • Data Transfert without special software
  • Two-detector Optical System
  • Range: 0-1000 NTU
  • Resolution: 0,01 on lowest range

Specifications
Measurement Method
Ratio turbidimetric determination using a primary nephelometric light
scatter signal (90°) to the transmitted light scatter signal
Regulatory
Meets EPA Method 180.1
Lamp Source Tungsten filament lamp
Range 0-1000 NTU
Accuracy ±2% of reading plus stray light from 0-1000 NTU
Repeatability ±1% of reading or 0.01 NTU, whichever is greater
Resolution 0.01 NTU on lowest range
Stray Light = 0.02 NTU
Signal Averaging Selectable on or off
Detector Silicon Photodiode
Reading Modes Normal (Push to Read)
Signal Averaging
Rapidly Settling Turbidity(TM)
Calibration Options Single step RapidCal(TM) for low-level regulatory reporting from 0-40 NTU
Full range calibration from 0-1000 NTU
Calibration to degrees of turbidity
Calibration Logger

Records the last 25 successful calibrations

Verification Logger Logs the last 250 successful verifications
Data Logger 500 records
Power Requirement AC 100-240 V, 50/60 Hz (with optional Power or USB+Power Module)
4 AA alkaline batteries
Rechargeable NiMH (with optional USB+Power Module)
Operating Conditions -40 to 60 °C (-40 to 140 °F), instrument only
Storage Conditions After 15 minutes of non-use
Interface Optional USB
Sample Required 15 mL (0.5 oz.)
Sample Cells Round cells, 60 x 25 mm (2.36 x 1 in.), borosilicate glass with screw caps
Dimensions 22.9 x 10.7 x 7.7 cm (9.0 x 4.2 x 3.0 in.)
Weight 530 g (1.17 lb) without batteries
620 g (1.37 lb) with four AA alkaline batteries
Meter Enclosure Rating IP67 (closed lid, battery and module compartment excluded)
Protection Class Power supply: Class II
Certification CE certified
Warranty 1 year