GENEQ > Environment > Flowmeters
| Hand Held Doppler (non-invasive) FLOWMETER |
| FEATURES |
|
| PRINCIPLE OF OPERATION |
The HFM Flowmeter is a non-invasive
device utilizing the frequency shift (Doppler effect) of a signal to measure
the flow velocities of liquids. The transmitters feed a high frequence
voltage to transmitter crystal of the transducer. The crystal converts
the voltage to a high frequency mechanical signal which is projected through
the wall of the pipe and into the liquid. The signals are reflected back
to a receiver crystal by the discontinuities in the fluid. The frequency
difference between the transmitted and received signal is directly proportional
to the fluid's flow velocity. In order to receive an adequate frequency
shift, the liquid must contain reflective particles such as air bubbles,
minute particles or other well-defined discontinuities. |
| APPLICATIONS |
The normal fields of application are
where conventional flowmeters do not work or create problems. Doppler
flowmeters are usually more cost-effective in that they do not vary in
price with an increase in pipe size. These instruments normally require
no change in the piping system and installation is fast and inexpensive. |
| WATER TREATMENT |
|
| CHEMICAL |
|
| PAPER |
|
| MINING |
|
| SPECIFICATIONS | |||
| HFM-1 | HFM-2 | HFM-4 | |
| Flow Range | 0.3 to 6 m/sec. - 0.5 to 20 ft/sec. | 0.5 to 20 ft/sec. and GPM | |
| Accuracy | Digital readout:
±2.0 L.S. Repeatability: ±0.2% (F.S. range) Linearity: ±0.5% (F.S. range) Resolution: 0.1 m/sec |
||
| Power | 5 VDC, supplied by internal nickel-cadmium batteries (or Charger below) | ||
| Transducer | Dual crystal hinged for ease of alignment | ||
| Operating Temperature | Electronics
-25°C to 85°C Transducer -40°C to 150°C |
||
| Readout | 12.5m LCD display with low battery indicator | ||
| Analog Output | N/A | 4 to 20 mA. DC | N/A |
| Size | 7.5" long x 4.25" wide x 2" deep | ||
| Weight | 1.8 kilograms /4 lbs. | ||
| Chargers | 2 styles available | ||
| Case | Padded nylon case holds flowmeter, transducer, interconnecting cables, couplant & charger. | ||
| CAT. NO. | DESCRIPTION |
| AUHFM-1000 | HFM-1 Flow Meter |
| AUHFM-2000 | HFM-2 Flow Meter with 4-20mA output |
| AUHFM-4000 | HFM-4 Flow Meter with range of 0.5 to 20 ft./sec and GPM |