The HI 8410 is an extended range panel
mounted Dissolved Oxygen (DO) controller
that is used to maintain and monitor the
concentration of DO in a wide range of
applications from wastewater treatment
to aquaculture. The HI 8410 uses a
Galvanic probe and typically requires less
maintenance than a Polarographic style
making it ideal for long term monitoring
of aeration basins, ponds, and lakes. The
set point for controlling the activation of
a relay is adjusted manually by the user.
An alarm relay is also manually adjustable
and is based upon a tolerance from the
programmed set point. Calibration is
single set point and can be done in zero
oxygen solution.
The dosage mode: Automatic, Continuous
ON or OFF and over dosage control by
setting the Overtime dosage trimmer. If
the dosing relay remains continuously
activated for more than the selected
dosing time, the alarm relay is activated,
the alarm LED will start blinking and the
dosing relay will be deactivated. A jumper
found on the rear panel can disable the
“Over time dosage” function.
“Automatic/Off/Manual” dosing selection
switch and LED on the front panel.
In Automatic mode all the relays are
controlled based on the measurement set
point and alarm values. In OFF mode the
dosing and alarm relays are always
deactivated. The dosing LED is OFF (as
relay status) and the ALARM LED is in
accordance with the instrument setpoint,
input reading, and ΔALARM.
In ON (Manual) mode the dosing relay is
always on. The alarm relay is still enabled.
If an alarm occurs the dosing relay
remains activated. If the over dosage time
exceeds the setting during manual mode,
the alarm relay remains activated.
The front panel is protected behind a
transparent splash-proof cover.
Features
Extended DO range to 50 mg/L (ppm)
Manual single point calibration
Selectable 0-20 or 4-20 mA output
Low maintenance Galvanic probe
Backlit LCD
Operational mode LED indicators
Specifications
Ordering information
The HI 8410 is supplied complete with mounting brackets and instructions.