Waterproof, Multi-Range, ATC Conductivity Meter
Model HI9033 & HI9034

hi9033.jpg (18377 octets)

HI 9033 and HI 9034 have been proven time and again to hold up under extended use in wet, humid, dusty and muddy conditions. HI 9033 utilizes a single probe to perform all conductivity readings in four ranges. This allows the user to measure samples from deionized water to brine without having to switch or recalibrate the probe.


HI 9034 measures total dissolved solids (TDS) in three ranges to offer the highest accuracy when performing measurements in diverse applications such as HVAC, waste water treatment and reverse osmosis. All three ranges can be activated at the touch of a button without having to change the conductivity probe.
Both meters perform measurements with ATC (automatic temperature compensation) which adjusts for the effects of temperature. These portable instruments also feature BEPS (Battery Error Prevention System) technology that alerts the user when low batteries could affect the readings.


FEATURES
One Probe for Many Ranges
The HI 76302W doesn’t need to be recalibrated or changed to measure in another range.
Waterproof
The HI 9033 & HI 9034 will resist the ingress of water & dust to protect its advanced circuitry, in industrial and field applications.
Four-Ring Method Excels
The proven 4-ring potentiometric method is so technologically advanced that it renders the 2-ring Amperometric method obsolete.

SPECIFICATIONS
HI9033 HI9034
Range 0.0 to 199.9 µS/cm; 0 to 1999 µS/cm;
0.00 to 19.99 mS/cm; 0.0 to 199.9 mS/cm
0.0 to 199.9 mg/L; 0 to 1999 mg/L;
0.00 to 19.99 g/L
Resolution 0.1 µS/cm; 1 µS/cm;
0.01 mS/cm; 0.1 mS/cm
0.1 mg/L; 1 mg/L; 0.01 g/L
Accuracy (@20°C/68°F) ±1% Full Scale (excluding probe error)
Typical EMC deviation ±1% Full Scale
Calibration Manual single point through trimmer
Temperature compensation Automatic from 10 to 40°C (50 to 104°F) with ß of 2% per degree °C
Probe HI 76302W with ATC 1m (3.3') screened cable (included)
Battery type / life 1x9V / 100 hours approx. of continuous use
Environment 0 to 50°C (32 to 122°F); RH 100%
Dimensions 196 x 80 x 60 cm (7.7 x 3.1 x 2.4")
Weight 425g (15oz.); Kit: 1.4kg (3.1 lb.)

CAT. NO. DESCRIPTION
HI9033 Supplied complete with HI 76302W interchangeable probe with 1 m (3.3’) screened cable, calibration screwdriver, 9V battery and instruction manual in a rugged carrying case.
HI9034 Supplied complete with HI 76302W interchangeable probe with 1 m (3.3’) screened cable, calibration screwdriver, 9V battery and instruction manual in a rugged carrying case.
HI 76302W Conductivity probe
HI 721317 Hard carrying case
HI 7030L 12880 µS/cm calibration solution (460 mL)
HI 7031L 1413 µS/cm calibration solution (460 mL)
HI 7032L 1382 mg/L calibration solution (460 mL)
HI 7033L 84 µS/cm calibration solution (460 mL)
HI 7034L 80000 µS/cm calibration solution (460 mL)
HI 7035L 111800 µS/cm calibration solution (460 mL)