GENEQ > Environnement > Turbidité > Chlore
| Turbidimètre et Chlorinomètre |
| Modèle HI93414 |
| HI98703 EPA Compliant
Turbidity Meter HI93414 EPA Compliant Turbidity and Chlorine Meter |
At startup, both meters display the
remaining battery level percentage and a low battery warning is also displayed
on the LCD to avoid unexpected battery failure. Each instrument is equipped
with backlight and the current time can be displayed on the LCD. |
| HI 98703 and
HI 93414 utilize an EPA compliant tungsten light source and state-of-the-art
optical system that allow consistently reliable and accurate measurements
of turbidity and free/total chlorine. Users will appreciate this accuracy
particularly at very low turbidity levels. HI 93414 meets the requirements
of EPA and Standard Methods for Turbidity and Colorimetric Chlorine Measurements. |
| Features |
• Tungsten light
source – EPA compliant turbidity measurement • High accuracy at low ranges (below 0.05 NTU) • Exclusive chlorine Cal Check™ calibration validation • Exclusive T.I.S. Tag Identification System • User replaceable light source • 2, 3 or 4 point turbidity calibration • USB and RS 232 connectivity • Backlit LCD • GLP capability • User friendly display with guidance codes • Auto shut-off • Battery percentage display • Continuous display of current time |
Turbidity
|
| Turbidity of water is
an optical property that causes light to be scattered rather than transmitted.
The scattering of light that passes through a liquid is primarily caused
by suspended solids. Higher turbidity means greater amounts of scattered
light. No solution will have zero turbidity because even the molecules
in a very pure fluid scatter light to a certain degree. The USEPA Method
180.1 specifies the key parameters for optical systems to measure drinking,
saline and surface water turbidity in a 0 to 40 NTU range, using the nephelometric
method. |
![]() |
The light beam that passes
through a sample is scattered in all directions. The intensity and pattern
of the scattered light is affected by many variables like wavelength of
the incident light, particle size, shape, refractive index and color.
The microprocessor of the instrument performs calculations using the signal
that reaches the two detectors to display the NTU value. The optical system
and measuring technique allow the compensation of lamp intensity fluctuations,
minimizing the need of frequent calibration. The lower detection limit
of a turbidimeter is determined by the so called “stray light”.
Stray light is the light detected by the sensors that is not caused by
light scattering from suspended particles. The optical system of these
instruments is designed to have very low stray light, providing accurate
results for low turbidity samples every time. |
| CAL CHECK™ Calibration Validation | |
When performing chlorine measurements,
you need to know that the instrument you are using is right on. With HANNA’s
exclusive CAL CHECK™ feature you can now rest assured. Simply insert
the factory calibrated standard of a known concentration and verify that
your instrument is accurate. With HANNA’s exclusive CAL CHECK™ validation function users are able to verify the performance of the instrument at any time. Taking just a few short steps, the validation procedure is user friendly and ensures that the meter is properly calibrated. Just use the exclusive HANNA ready-made, NiST traceable standards to verify the performance of the instrument and recalibrate if necessary. All instruments are factory calibrated and the electronic and optical design minimizes the need for frequent calibration. |
![]() |
| SPECIFICATIONS | |
| Turbidity (HI 93414 & HI 98703) | |
| Range | 0.00 to 9.99; 10.0 to 99.9 and 100 to 1000 NTU |
| Range Selection | Automatic |
| Resolution | 0.01 NTU from 0.00 to 9.99 NTU; 0.1 NTU
from 10.0 to 99.9 NTU; 1 NTU from 100 to 1000 NTU |
| Accuracy | ±2% of reading plus 0.02 NTU |
| Repeatability | ±1% of reading or 0.02 NTU, whichever is greater |
| Stray Light | < 0.02 NTU |
| Typical EMC Deviation | ±0.05 NTU |
| Light Detector | Silicon Photocell |
| Method | Ratio Nephelometric Method (90°),
ratio of scattered and transmitted light; Adaptation of the USEPA Method 180.1 and Standard Method 2130 B. |
| Measuring mode | Normal, Average, Continuous |
| Turbidity Standards | 1, 10, 100 and 750 NTU |
| Calibration | Two, three or four-point calibration |
| Free and Total Chlorine (HI 93414 only) | |
| Range | Free Cl2 0.00 to 5.00 mg/L; Total Cl2 0.00 to 5.00 mg/L |
| Resolution | 0.01 mg/L from 0.00 to 3.50 mg/L; 0.10 above 3.50 mg/L |
| Accuracy | ±0.02 mg/L @ 1.00 mg/L |
| Typical EMC Deviation | ±0.02 mg/L |
| Detector | Silicon photocell with 525 nm narrow band interference filters |
| Method | Adaptation of the USEPA Method 330.5
and Standard Method 4500- Cl G. The reaction between chlorine and DPD reagent causes a pink tint in the sample |
| Standards | 1 mg/L free chlorine, 1 mg/L total chlorine |
| Calibration | One-point calibration |
| Common Specifications | |
| Light Source | Tungsten filament lamp |
| Lamp life | Greater than 100,000 readings |
| Display | 60 x 90 mm LCD with back light |
| LOG Memory | 200 records |
| Serial Interface | USB 1.1 or RS 232 |
| Environment | To 50°C (122°F); max 95% RH non-condensing |
| Power Supply | (4) 1.5V AA alkaline batteries or AC adapter |
| Auto-off | After 15 minutes of non-use |
| Dimensions | 224 x 87 x 77 mm (8.8 x 3.4 x 3.0”) |
| Weight | 512 g (18 oz.) |
| ACCESSORIES | |
| HI 710005 Voltage adapter | HI 92000 Windows® compatible software |
| HI 731318 Tissue for wiping cuvets (4 pcs) | HI 920005 5 tag holders with tags |
| HI 731331 Glass cuvets (4 pcs) | HI 920011 5 to 9 pin RS232 connection cable |
| HI 740027P 1.5V AA battery (12 pcs) | HI 93703-50 Cuvet cleaning solution (230 mL) |
![]() |
ORDERING
INFORMATION HI 93414 Portable Turbidity and Free/Total Chlorine & HI 98703 portable turbidity meters are supplied with 1 - Five sample cuvets and caps 2 - Three calibration cuvets for turbidimeter 3 - Two calibration cuvets for colorimeter* 4 - Reagent powder packets for free & total chlorine* 5 - Silicone oil 6 - Five tag holders with tags (HI 920005) 7 - Scissors 8 - Batteries (4 pcs.) 9 - AC adapter 10 - Rugged carrying case Instruction manual Quick reference guide Instrument quality certificate |
![]() |