GENEQ > Environnement > Turbidité  > Chlore

Turbidimètre et Chlorinomètre
Modèle HI93414



HI98703 EPA Compliant Turbidity Meter
HI93414 EPA Compliant Turbidity
and Chlorine Meter
At startup, both meters display the remaining battery level percentage and a low battery warning is also displayed on the LCD to avoid unexpected battery failure. Each instrument is equipped with backlight and the current time can be displayed on the LCD.

HI 98703 and HI 93414 utilize an EPA compliant tungsten light source and state-of-the-art optical system that allow consistently reliable and accurate measurements of turbidity and free/total chlorine. Users will appreciate this accuracy particularly at very low turbidity levels. HI 93414 meets the requirements of EPA and Standard Methods for Turbidity and Colorimetric Chlorine Measurements.
Turbidity is measured up to 1000 NTU and free or total chlorine measurements can be made in the 0.00 to 5.00 mg/L (ppm) range. To meet EPA reporting requirements these instruments have an EPA compliance reading mode.
Depending on the measured sample and required accuracy, users can select normal measurement, continuous measurement or signal averaging.
Both models feature Hanna’s new Tag Identification System (T.I.S.) which allows users to record the time and place at unlimited specific measurement locations.
Logged data can be downloaded through either USB or RS 232.
Reliable performance of the instrument is validated using HANNA’s exclusive CAL CHECK™ system and our ready-made, NIST traceable chlorine standards. For turbidity, a two, three or four-point calibration is available using pre-set or user defined standards. Frequent calibration isn’t necessary as the system compensates for variations in intensity of the light source. The instruments’ Good Laboratory Practice functions allow traceability of calibration conditions including the time, date and last calibration points.


Features

• Tungsten light source – EPA compliant turbidity measurement
• High accuracy at low ranges (below 0.05 NTU)
• Exclusive chlorine Cal Check™ calibration validation
• Exclusive T.I.S. Tag Identification System
• User replaceable light source
• 2, 3 or 4 point turbidity calibration
• USB and RS 232 connectivity
• Backlit LCD
• GLP capability
• User friendly display with guidance codes
• Auto shut-off
• Battery percentage display
• Continuous display of current time

Turbidity

Turbidity of water is an optical property that causes light to be scattered rather than transmitted. The scattering of light that passes through a liquid is primarily caused by suspended solids. Higher turbidity means greater amounts of scattered light. No solution will have zero turbidity because even the molecules in a very pure fluid scatter light to a certain degree. The USEPA Method 180.1 specifies the key parameters for optical systems to measure drinking, saline and surface water turbidity in a 0 to 40 NTU range, using the nephelometric method.
HI 98703 and HI 93414 are designed to meet or exceed the criteria specified by the USEPA Method 180.1 and Standard Method 2130 B. Model HI 93414 is based on a state-of-the-art optical system that guarantees both high performance and reliable results: this optical system includes a tungsten filament lamp, a scattered light detector and a transmitted light detector.

The light beam that passes through a sample is scattered in all directions. The intensity and pattern of the scattered light is affected by many variables like wavelength of the incident light, particle size, shape, refractive index and color. The microprocessor of the instrument performs calculations using the signal that reaches the two detectors to display the NTU value. The optical system and measuring technique allow the compensation of lamp intensity fluctuations, minimizing the need of frequent calibration. The lower detection limit of a turbidimeter is determined by the so called “stray light”. Stray light is the light detected by the sensors that is not caused by light scattering from suspended particles. The optical system of these instruments is designed to have very low stray light, providing accurate results for low turbidity samples every time.

CAL CHECK™ Calibration Validation
When performing chlorine measurements, you need to know that the instrument you are using is right on. With HANNA’s exclusive CAL CHECK™ feature you can now rest assured. Simply insert the factory calibrated standard of a known concentration and verify that your instrument is accurate.
With HANNA’s exclusive CAL CHECK™ validation function users are able to verify the performance of the instrument at any time. Taking just a few short steps, the validation procedure is user friendly and ensures that the meter is properly calibrated. Just use the exclusive HANNA ready-made, NiST traceable standards to verify the performance of the instrument and recalibrate if necessary. All instruments are factory calibrated and the electronic and optical design minimizes the need for frequent calibration.

SPECIFICATIONS
Turbidity (HI 93414 & HI 98703)
Range 0.00 to 9.99; 10.0 to 99.9 and 100 to 1000 NTU
Range Selection Automatic
Resolution 0.01 NTU from 0.00 to 9.99 NTU; 0.1 NTU from 10.0 to 99.9 NTU;
1 NTU from 100 to 1000 NTU
Accuracy ±2% of reading plus 0.02 NTU
Repeatability ±1% of reading or 0.02 NTU, whichever is greater
Stray Light < 0.02 NTU
Typical EMC Deviation ±0.05 NTU
Light Detector Silicon Photocell
Method Ratio Nephelometric Method (90°), ratio of scattered and transmitted light;
Adaptation of the USEPA Method 180.1 and Standard Method 2130 B.
Measuring mode Normal, Average, Continuous
Turbidity Standards 1, 10, 100 and 750 NTU
Calibration Two, three or four-point calibration
Free and Total Chlorine (HI 93414 only)
Range Free Cl2 0.00 to 5.00 mg/L; Total Cl2 0.00 to 5.00 mg/L
Resolution 0.01 mg/L from 0.00 to 3.50 mg/L; 0.10 above 3.50 mg/L
Accuracy ±0.02 mg/L @ 1.00 mg/L
Typical EMC Deviation ±0.02 mg/L
Detector Silicon photocell with 525 nm narrow band interference filters
Method Adaptation of the USEPA Method 330.5 and Standard Method 4500-
Cl G. The reaction between chlorine and DPD reagent causes a pink
tint in the sample
Standards 1 mg/L free chlorine, 1 mg/L total chlorine
Calibration One-point calibration
Common Specifications
Light Source Tungsten filament lamp
Lamp life Greater than 100,000 readings
Display 60 x 90 mm LCD with back light
LOG Memory 200 records
Serial Interface USB 1.1 or RS 232
Environment To 50°C (122°F); max 95% RH non-condensing
Power Supply (4) 1.5V AA alkaline batteries or AC adapter
Auto-off After 15 minutes of non-use
Dimensions 224 x 87 x 77 mm (8.8 x 3.4 x 3.0”)
Weight 512 g (18 oz.)

ACCESSORIES
HI 710005 Voltage adapter HI 92000 Windows® compatible software
HI 731318 Tissue for wiping cuvets (4 pcs) HI 920005 5 tag holders with tags
HI 731331 Glass cuvets (4 pcs) HI 920011 5 to 9 pin RS232 connection cable
HI 740027P 1.5V AA battery (12 pcs) HI 93703-50 Cuvet cleaning solution (230 mL)

ORDERING INFORMATION
HI 93414 Portable Turbidity and Free/Total Chlorine & HI 98703
portable turbidity meters are supplied with
1 - Five sample cuvets and caps
2 - Three calibration cuvets for turbidimeter
3 - Two calibration cuvets for colorimeter*
4 - Reagent powder packets for free & total chlorine*
5 - Silicone oil
6 - Five tag holders with tags (HI 920005)
7 - Scissors
8 - Batteries (4 pcs.)
9 - AC adapter
10 - Rugged carrying case
Instruction manual
Quick reference guide
Instrument quality certificate