GENEQ > Environment > Conductivity
| Conductivity/TDS/°C/NaCl Meter |
| Model HI9835 |
The HI 9835 combines the more important parameters you need for measuring conductivity. Now you can sample Electro-Conductivity (EC), Total Dissolved Solids (TDS), sodium chloride (NaCl) and temperature all at once. Hanna's probe design innovation uses 4 rings enhanced with platinum for greater stability while increasing the range of measurable concentrations and temperature. You can choose between automatic and manual temperature compensation to facilitate your operating requirements. In addition, you can also select from a range of TDS factors from 0.40 to 0.80 and a range of temperature coefficients 0.00 to 6.00% for greater consistency and reproducibility. The housing is completely waterproof and built to survive the harsh conditions of field use. For extended time studies, this meter can be connected to the available 12 VDC power supply. |
| SPECIFICATIONS | ||||
|
EC
|
TDS
|
NaCl
|
Temp.
|
|
| Range (autoranging) |
0.00 to 29.99 µS/cm
30.0 to 299.9 µS/cm 300 to 2999 µS/cm 3.00 to 29.99 mS/cm 30.0 to 200.0 mS/cm Up to 500.0 mS/cm |
0.00 to 14.99 ppm
15.0 to 149.9 ppm 150 to 1499 ppm 1.50 to 14.99 ppt 15.0 to 100.0 ppt Up to 400.0 ppt |
0.0 to 400.0%
|
0.0 to 60.0°C
|
| Resolution |
0.01 µS/cm (from 0.00 to 29.99
µS/cm)
0.1 µS/cm (from 30.0 to 299.9 µS/cm) 1 µS/cm (from 300 to 2999 µS/cm) 0.01 mS/cm (from 3.00 to 29.99 mS/cm) 0.1 mS/cm (over 30.0 mS/cm) |
0.01 ppm (from 0.00 to 14.99 ppm)
0.1 ppm (from 15.0 to 149.9 ppm) 1 ppm (from 150 to 1499 ppm) 0.01 ppt (from 1.50 to 14.99 ppt) 0.1 ppt (over 15.0 ppt) |
0.1%
|
0.1°C
|
| Accuracy |
±1% of reading ± (0.05
µS/cm or 1 digit, whichever is greater)
|
±1% of reading ± (0.03
ppm or 1 digit, whichever is greater)
|
±1% of reading
|
±0.4°C
|
| Typical EMC Deviation |
±1% of reading
|
±1% of reading
|
±1% of reading
|
±0.1°C
|
| EC Calibration |
1 point with 6 memorized buffers
|
|||
| NaCl Calibration |
1 point with HI 7037 buffer (optional)
|
|||
| Temperature Calibration |
2 points at 0 and 50°C (plus ±1°C
adjustment)
|
|||
| Temperature Compensation |
Automatic or manual from 0 to 60°C
(can be disabled to measure actual conductivity)
|
|||
| Temperature Coefficient |
0.00 to 6.00%/°C (for EC and TDS
only) Default value is 1.90%/°C
|
|||
| TDS Factor |
0.40 to 0.80 (default value is 0.50)
|
|||
| Probe |
HI 76309 4-ring probe, K=1 nominal and
built-in temperature sensor (included)
|
|||
| Auto Off |
After 5 minutes (can be disabled)
|
|||
| Power Supply |
4 x 1.5V AA alkaline type batteries (included),
or 12VDC adapter
|
|||
| Environment |
0 to 50°C; 100% RH
|
|||
| Dimensions |
196 x 80 x 60 mm (7.7 x 3.1 x 2.4)
|
|||
| Weight |
500 g (18 oz)
|
|||
| CAT. NO. | DESCRIPTION |
| HI9835 | Supplied complete with HI 76309 4-ring conductivity probe with 1 m (3.3) screened cable, 4 x 1.5V AA batteries and instruction manual in a rugged case. |
| HI 76309 | Conductivity probe |
| HI 721317 | Hard carrying case |
| HI 710005 | 110V to 12VDC adapter |
| HI 710006 | 220V to 12VDC adapter |
| HI 7030L | 12880 µS/cm calibration solution (460 mL) |
| HI 7031L | 1413 µS/cm calibration solution (460 mL) |
| HI 7033L | 84 µS/cm calibration solution (460 mL) |
| HI 7037L | NaCl calibration solution (460 mL) |