| SPECIFICATIONS |
| Performance |
| Water Activity |
| Measurement range |
0 to 1 aw |
| Accuracy (including nonlinearity,
hysteresis and repeatability) when calibrated agaisnt salt solutions
(ASTM E104-85) |
±0.02 (0 to 0.9)
±0.03 (0.9 to 1.0) |
| Accuracy (including nonlinearity,
hysteresis and repeatability) maximem achievable accuracy when calibrated
against high quality, certified humidity standards |
±0.01 (0 to 0.9)
±0.02 (0.9 to 1.0) |
| Response time (90%) |
10 min. (at 20°C)
in still oil with stainless steel filter |
| Sensor |
Vaisala HUMICAP |
| Recommended recalibration
interval |
1 year |
| Temperature |
| Measurement range |
-40 to 100°C (-40
to 212°F) |
| Typical accuracy at 20°C |
±0.2°C (±0.36°F) |
| Typical temp. dependence
of electronics |
±0.005°C/°C
(±0.005°F/°F) |
| Sensor |
Pt 100 IEC 751 1/3 class
B |
| Typical long-term stability |
better than 0.001 aw/year |
| Operating Environment |
| Probe |
| Operating temp. for electronics |
-40 to 60°C (-40
to 140°F) |
| Operating pressure range |
Max. 20 bar
Max. 10 bar (during installation through ball valve) |
| Oil flow range |
Max. 1 m/s |
| Indicator |
| Operating temp. range |
-10 to 40°C (14 to
104°F) |
| Operating humidity range |
non-condensing |
| Electromagnetic compatibility |
| Complies with |
EMC standard EN61326-1:1997
+ Am1:1998, Electrical equipment fr measurement, control and laboratory
use - EMC requirements: Portable test and measurement. |
| Inputs and Outputs |
| Power supply |
Rechargeable NiMH battery
pack with AC-POWER or 4 x AA-size alkalines, type IEC LR6 |
| Battery operation time
continuous use |
48 hours typical at 20°C
(68°F) |
| Battery operation time
datalogging use |
Up to 31 days |
| Menu languages |
English, French, Spanish,
German, Sweidish, Finninsh |
| Display |
LCD with backlight, graphic
trend display of any parameter character hight up to 16mm |
| Analog output |
0 to 1 VDC |
| Output resolution |
0.6 mV |
| Data interface |
RS232 |
| Datalogging capacity |
2700 points |
| Alarm |
Audible alarm function |
| Mechanics |
| Probe |
| Housing classification |
IP65 (NEMA 4) |
| Housing material |
ABS/PC blend) |
| Probe material |
Stainless steel (AISI316L) |
| Cable length between
probe and indicator |
1.9m (10m extension available) |
| Weight |
506g |
| Indicator |
| Housing classification |
IP54 |
| Weight |
400g |
| Probe inputs |
1 or 2 |
| CAT.
NO. |
DESCRIPTION |
| VAMM70-000 |
Handheld moisture and temperature meter
for spot-checking in oil |
| Probes |
| VAHM75-000 |
Relative humidity measurement probe model
HM75 |
| VAHM76-000 |
Relative humidity measurement probe model
HM76 |
| VAHM77-000 |
Relative humidity measurement probe model
HM77 |
| VADMP74A/B |
Dewpoint measurement probes |
| Options and accessories |
| VAMI70CASE2 |
Carrying case |
| VAHMP228BVS |
Ball valve set (incl. fitting body and
blanking plug) |
| VA213017SP |
Probe cable extension, 10m |
| VA27159ZZ |
Transmitter connection cable for HMP228 |
| VADRW216050 |
Transmitter connection cable for MMT318 |
| VA27168ZZ- |
Analog output cable |
| VAMI70LINK |
Windows software, inclides PC connection
cable |
| VAHM47453SP |
Sensor protection |