GENEQ > Meteorology > Dataloggers

Humidity-Temperature Data Logger
Model HygroLog

humidity-temperature_data_logger.jpg (11045 octets)

FEATURES
  • Advanced sensor technology with microprocessor and ASIC provides accuracy better than ±1.5%RH and ±0.3°C
  • Fully programmable with alarm function
  • Memory capacity of 5,000 humidity-temperature data points
  • Up to one year operation without battery replacement
  • Powerful, user friendly Windows based software
  • Probe can be separated from electronics for measurements in the range of -40...85° (-40...185°F)
  • HygroClip plug-in probe is supported by our worldwide exchange program and is quickly replaced with no loss of accuracy and no calibration

SPECIFICATIONS
 
HygroLog
HygroLog-D
Plug-in probe
HygroClip S (included)
Humidity sensor
ROTRONIC HYGROMER C94 (capacitive)
Temperature sensor
Pt100 1/3 DIN RTD
Operating Range - Logger
0...100% RH
-10...50°C with alkaline batteries
-20...70°C with lithium batteries
0...100% RH
-10...50°C (limited by LC display)
Accuracy at 23°C
Better than ±1.5% and ±1.5%RH and ±0.2°C / ±0.4°F
Repeatability
<0.5%RH and 0.1°C
Memory
5,000 x humidity and temperature data points
Log Internal
15 sec to 120 min
Batteries
3 x 1.5 V
Battery life (5min. log interval)
about 1 year
about 6 months
External Power
8...20 VDC (requires HYGROLOG-IPH holder and AC adapter)
PC requirements
Microsoft Windows 95/98 or NT, MSIE 4.0 Browser or higer 90 MHz Pentium, 32 MB RAM, VGA monitor 800x600, 256 colors or higher
Case material
ABS
Protection Grade
IP65
Dimensions
140 x 85 x 25 mm (5.51 x 3.35 x 0.98")
Weight
200g (0.44 lb)

SOFTWARE AND PROGRAMMING
Continuous (Loop) Recording Data recording begins as soon as the HygroClip is plugged in. When the memory is full, the oldest data is erased as new data is added.
Start/Stop Recording Data recording begins at a specific times and ends when the memory is full.
Alarm Values Alarms values can be programmed and shown on a chart. The display of the HygroLog-D provides an indication when any alarm value has been exceeded.
HW3 Software The HW3 software is used to program and read the Hygrolog. The data can be displayed as a table or as a graph. If so desired, HW3 can also be used to calibrate the HygroClip.

ACCESSORIES
HYGRODATA-HL HW3 software and data cable
HYGROLOG-IPH Wall-mount holder, can be used with or without AC adapter
HYGROLOG-SP Shock protection for HygroLog/D
MOK-xx-DAT05 Extension cable for the HygroClip
xx = 01 (1m/3ft) or 02 (2m/6ft) or 05 (5m/15ft maximum length)
HYGROCLIP-R Replacement plug-in probe 100% interchangeable without calibration
ER-15 Calibration device for HygroCLip
EAxx-SCS Certified humidity standards )box of 5)
xx = 0, 5, 10, 20, 35, 50, 65, 80, 95%RH
MOK-02-WIN Calibration cable for HygroClip (plugs into the COM port of a PC)