GENEQ > Meteorology > Dataloggers
Humidity-Temperature Data Logger
Model HygroLog

| FEATURES |
|
| SPECIFICATIONS | ||
|
HygroLog
|
HygroLog-D
|
|
| Plug-in probe |
HygroClip S (included)
|
|
| Humidity sensor |
ROTRONIC HYGROMER C94 (capacitive)
|
|
| Temperature sensor |
Pt100 1/3 DIN RTD
|
|
| Operating Range - Logger |
0...100% RH
-10...50°C with alkaline batteries -20...70°C with lithium batteries |
0...100% RH
-10...50°C (limited by LC display) |
| Accuracy at 23°C |
Better than ±1.5% and ±1.5%RH
and ±0.2°C / ±0.4°F
|
|
| Repeatability |
<0.5%RH and 0.1°C
|
|
| Memory |
5,000 x humidity and temperature
data points
|
|
| Log Internal |
15 sec to 120 min
|
|
| Batteries |
3 x 1.5 V
|
|
| Battery life (5min. log interval) |
about 1 year
|
about 6 months
|
| External Power |
8...20 VDC (requires HYGROLOG-IPH
holder and AC adapter)
|
|
| PC requirements |
Microsoft Windows 95/98 or NT,
MSIE 4.0 Browser or higer 90 MHz Pentium, 32 MB RAM, VGA monitor 800x600,
256 colors or higher
|
|
| Case material |
ABS
|
|
| Protection Grade |
IP65
|
|
| Dimensions |
140 x 85 x 25 mm (5.51 x 3.35
x 0.98")
|
|
| Weight |
200g (0.44 lb)
|
|
| SOFTWARE AND PROGRAMMING | |
| Continuous (Loop) Recording | Data recording begins as soon as the HygroClip is plugged in. When the memory is full, the oldest data is erased as new data is added. |
| Start/Stop Recording | Data recording begins at a specific times and ends when the memory is full. |
| Alarm Values | Alarms values can be programmed and shown on a chart. The display of the HygroLog-D provides an indication when any alarm value has been exceeded. |
| HW3 Software | The HW3 software is used to program and read the Hygrolog. The data can be displayed as a table or as a graph. If so desired, HW3 can also be used to calibrate the HygroClip. |
| ACCESSORIES | |
| HYGRODATA-HL | HW3 software and data cable |
| HYGROLOG-IPH | Wall-mount holder, can be used with or without AC adapter |
| HYGROLOG-SP | Shock protection for HygroLog/D |
| MOK-xx-DAT05 | Extension cable for the
HygroClip xx = 01 (1m/3ft) or 02 (2m/6ft) or 05 (5m/15ft maximum length) |
| HYGROCLIP-R | Replacement plug-in probe 100% interchangeable without calibration |
| ER-15 | Calibration device for HygroCLip |
| EAxx-SCS | Certified humidity standards
)box of 5) xx = 0, 5, 10, 20, 35, 50, 65, 80, 95%RH |
| MOK-02-WIN | Calibration cable for HygroClip (plugs into the COM port of a PC) |