GENEQ > Meteorology > Transmitters
| Intrinsically Safe Humidity and Temperature Transmitters |
| HMT360 Series |
|
HUMICAP humidity and temperature transmitter series HMT360
are the ideal solution for measuring humidity in hazardous areas. They operate
safely and reliably even in the most hazardous classifictions. The HMT360
transmitters' proven performance and technology conform with rigous international
standards.
|
| FEATURES & BENEFITS |
|
|
| INTRINSICALLY SAFE |
| The entire HMT360 transmitter can be installed directly in explosive areas. It can withstand continuous exposure to potentially explosive environments that contain flammable gases. |
| CUSTOMIZED CONFIGURATION |
| Due to the microprocessor based electronics, options and accessories, the HMT360 series is truly flexible. Customers may specify the transmitter configuration when ordering the instrument, however they may change the configuration in the field. |
| FIVE INTERCHANGEABLE PROBES | |
| The HMT360 offers five probe options for various applications | |
| HMP361 | Wall mount |
| HMP363 | Confined spaces |
| HMP364 | High pressure |
| HMP365 | High temperature |
| HMP368 | Pressurized pipelines |
|
The interchangeable probes enable fast and easy removal
or re-installation when required. Calibration, for example, is easy to perform
due to the modular structure. All calibration coefficients are included
in the probe unit itself, which means tht probes can be switched between
transmitter bodies without losing the accuracy.
|
|
| OPTIMIZED SENSORS |
| In addition to standard HUMICAP Sensor, also application specific, very chemically durable sensor is available. |
| LONG-TERM SOLUTION |
| The HMT360 transmitters are an investment: their rugged design, combined with trouble-free operation, ensure a long-term solution for monitoring humidity and dewpoint in explosive environments. |
| SPECIFICATIONS | |
| Measured Variables | |
| Relative humidity | |
| Measurement range | 0 to 100% |
| Accuracy (including nonlinearity, hysteresis and repeatability) Max. achievable accuracy when calibrated against high quality, vertified humidity standards | ±1% RH (0 to 90%RH) ±2% RH (90 to 100%RH) |
| Accuracy (including nonlinearity, hysteresis and repeatability) When calibrated against salt solutions (ASTM E104-85) | ±2% RH (0 to 90%RH) ±3% RH (90 to 100%RH) |
| Response time (90%) at 20°C | 15 seconds (in still air with sintered filter) |
| Sensors |
|
| Temperature | |
| Measurement range | -40 to 180°C (-40 to 356°F) depends on selected probe |
| Typical accuracy of electronics at 20°C | ±0.1°C (±0.18°F) |
| Typical temperature dependence of electronics | 0.005°C/°C (0.005°F/°F) |
| Sensor | Pt 1000 RTD 1/3 Clas B IEC 751 |
| Calculated variables (typical ranges) | |
| With HMP361 probe | |
| Dewpoint temperature | -20 to 60°C (-4 to 140°F) |
| Mixing ratio | 0 to 160 g/kg d.a. |
| Absolute humidity | 0 to 160g/m3 |
| Wet bulb temperature | 0 to 100°C (32 to 212°F) |
| With HMP363, HMP364, HMP365 and HMP368 Probes | |
| Dewpoint temperature | -20 to 100°C (-4 to 212°F) |
| Mixing ratio | 0 to 160 g/kg d.a. |
| Absolute humidity | 0 to 500 g/m3 |
| Wet bulb temperature | 0 to 100°C (32 to 212°F) |
| The accuracies of these calculated variables are limited by the accuracies of the measured variables on which they are based, namely RH and Temperature. The RH and Temperature accuracies are stated above under Measured Variables. | |
| Outputs | |
| Two analog outputs | Two wire 4 to 20 mA (on standard, one optional) |
| Typical accuracy of analog outputs at 20°C | ±0.05% full scale |
| Typical temperature dependence of analog outputs | 0.005%/°C (0.005%°F) full scale |
| Analog outputs | Connection via safety barriers |
| RS232C serial output for service use | Connector type RJ45 |
| Classification with current outputs | |
| Europe / CENELEC (PTB) | |
| EU (94/9/EC, ATEX100a) | II1G, EEx ia IIC T4 PTB 00 ATEX 2112 X |
| Safety factors | Ui=28 V, Ii=100mA, Pi=0.7W Ci=1 nF, Li=0 H |
| Environmental specifications | Tamb: -20 to 60°C (-4 to
140°F) Pamb: 0.8 to 1.1 bar |
| USA (FM) | |
| Classes I, II, III, Division 1, Groups
A-G and Division 2 Groups A-S-F and G FM Project ID: 3010615 |
|
| Safety factors | Vmax=28 VDC, Imax=100mA, Ci=1nF, Pi=0.7W, Li=0, Tamb = 60°C (140°F), T5 |
| Japan (TIIS) | |
| Ex ia IIC T4 Code number: C15354 |
|
| Safety factors | Ui=28 VDC, Ii=100 mA, Ci=1nF, Pi=0.7W, Li=0, Tamb = 60°C (140°F) |
| Australia (TestSafe) | |
| Ex ia IIC T5 IP65 Certificate no: Ex AUS Ex 3738X |
|
| Safety factors | Ui=28 V, Ii=100mA, Pi=0.7W, Ci=1 nF, Li=0 mH |
| Canada (CSA) | |
|
|
|
| Safety factors | Tamb = 60°C, T4, Intrinsically safe when connected as per installation drawing DRW213478 |
| General | |
| Operating voltage | 12 to 28V |
| Operating voltage with serial port (service mode) | 15 to 28V |
| Connections | Screw terminals, 0.33 to 2.0mm2 Wires (AWG 14-22) |
| Cable bushing | Pg11 (5 to 12mm) |
| Conduit fitting | Pg11/NPT 1/2 in.-14 |
| Operating temp. range for electronics | -40 to 60°C (-40 to 140°F) |
| Operating temp. range for electronics with display | -20 to 60°C (-4 to 140°C) |
| Storage temperature range | -40 to 70°C (-40 to 158°F) |
| Housing material | G-AlSi10Mg (DIN 1725) |
| Housing classification | IP65 (NEMA 4) |
| Housing weight | 950g |
| Display | Two-line LCD |
| Character Size | 1st line/2nd line) 12mm/10mm |
| Calculated output variables | Dewpoint temperature, mixing ratio, absolute humidity, wet bulb temperature |
| Additional analog output | 4 to 20 mA |
| Complies with | EMC standard EN61326-1:1997 + Am1:1998 Industrial environment |
| Note: IEC 1000-4-5 complies only when using
external Exi approved surge arrester on safe area |
|
| CAT. NO. | DESCRIPTION |
| VAHMT360-0 | Intrinsically safe humidity and temperature transmitter |
| Interchangeable Probes | |
| VAHMP361-0 | Interchageable probe for wall mounting |
| VAHMP363-0 | Interchangeable probe for confined spaces |
| VAHMP364-0 | Interchangeable probe for high pressure |
| VAHMP365-0 | Interchangeable probe for high temperature |
| VAHMP368-0 | Interchangeable probe for pressurizd pipelines |
| Options and Accessories | |
| VAHMP233FAH | Duct mounting installation kit for HMP363) |
| VAHMP235FA | Installation flange for HMP365 (aluminium) |
| VAHMP235FS | Installation flange for HMP365 (stainless steel) |
| VADMP248BVS | Ball valve set for HMP368. Pressure range at 20°C: 0 to 40 bar (Durig installation max. 10 bar) |
| VA25905ZZ- | Serial interface cable for PC connectors RJ45-D9 female |