GENEQ > Biotechnology > HPLC
Isocratic HPLC Systems
Models DLC-20 and DLC-20P
Although not designed to replace the more expensive
and larger systems , the DLC-20 and 20P, with their small size , price, and
versatility are an ideal answer for routine QC analysis. Several DLC-20/20P
can be dedicated to methods much less expensively than buying one larger and
more expensive gradient system, particularly since most QC procedures are isocratic
in nature.
The DLC-20 can be obtained in PEEK construction for bio- compatibility or in
SS at the same price. The DLC-20P is available only in SS construction. Similar
to the DLC-10CP/11CP the DLC-20P can operate to 5000 PSI with the standard 10ml
pump head (SS). A 5.0ml pump head (SS) is an additional accessory. All controls
are conveniently located on the front of the instrument. There are two LED displays.
One is dedicated to the pump while the other displays detector absorbance and
wavelength information. On the DLC-20P the pump display also indicates system
operating pressure. The rear panel has the analog voltage outputs for data acquisition
and outputs for contact closures permitting remote control of absorbance and
wavelength.
The DLC-20 comes standard with a two stage valve and 20µl loop (either
PEEK or SS). The 20P adds a position sensing valve.
| SPECIFICATIONS | |
| Flowrate | 0.01 to 9.99 mL/min |
| Pressure | 3000PSI / 5000PSI |
| Unswept Volume | <30µL |
| Display(s) | AU, Wavelength, P |
| Materials wetted | PEEK or SS |
| Accuracy Pump | 2% |
| Precision Pump | 0.2%RSD |
| Wavelength Accuracy | 2nm |
| Wavelength Precision | 1nm |
| Lamp | Deuterium |
| Flowcells | |
| Analytical PEEK, SS | 7.0mm Path; 10µL Vol |
| Semi-Prep PEEK, SS* | 2.5mm Path; 44µL Vol |
| Preparative PEEK | 2.0mm Path; 4µL Vol |
| Microbore, PEEK, SS | 5.0mm Path, 1.5 mL Vol* |
| Linearity | >2% |
| Drift | <2.5 x 10(-4) |
| Noise | <2.5 x 10(-4) |
| Recorder Output | 10.0mV |
| Integrater Output | 1.0V-2.0V (2.0A) |
| Pump Output | RS-232 |
| Power | 115, 220-240VAC |
| Dimensions | 7.1Hx7.8Wx16D in. |
| Weight | 40lbs (19.15kg) |
| * Fixed Wavelength Only | |
| CAT. NO. | DESCRIPTION |
| DLC-20 | Integrated isocratic system with variable wavelength UV-VIS detector. Includes analytical flow cell at no additional charge. |
| DLC-20P | The high pression version of DLC-20; operates at up to 5000 psi. Includes built-in pressure transducer, font panel display and high & low pressure alarms. |