The Series 700 family of submersible pressure transducers is
specifically designed to meet the rigorous environments encountered in liquid
level measurement and control. It can be configured to perform to specifications
under the most adverse, reactive conditions.
These transducers incorporate the latest advancements in piezoresistive pressure sensing technology. A stability-enhancing charged "Field Shield" is vapor deposited directly to the pressure cell. A welded 316 Stainless Steel diaphragm, with a spring rate ratio of 1000:1 with the piezoresistive pressure cell, is used for contact with the media. The transducer housing is an all-welded design, constructed of corrosion resistant 316 SST. Titanium housing with hastelloy/platinum sensor combination is offered for extremely corrosive media. The Series 700 also features state-of-the-art, surface mount internal signal conditioning which provides a power supply rejection of 0.001% and either a 4-20 mAdc, 0-5 Vdc, or mVdc process signal. Approvals to FM, CSA, and UL are available for Class I, Div 1, Groups A, B, C and D, and Class II, Div 1, Groups E, F and G, and Class III, Div 1 hazardous locations. These instruments also meet CE approval according to EN-50081-2 and EN-50082-2. Hazardous locations installation must be to local and national electrical codes and installed with an approved electrical barrier, such as manufactured by R.G. Stahl, Inc. Each transducer is shipped with a barrier that prevents moisture from entering the cable vent tube and with a traceable calibration card. The card specifies input/output conditions and actual data recorded during manufacture. Optional calibration is available when additional performance characteristics are required. All units are repairable and have low power requirements. |
|
SPECIFICATIONS
|
|||
| 730..... 720..... 710..... 700 | Units | Comments | |
| Pressure Ranges | |||
| Pressure Ranges | 0-5 through 0-300 0-35 through 0-21000 |
psi kPa |
any intermediate ranges available |
| Proof Pressure | 1.5 | x F.S. | |
| Burst Pressure | 2.0 | x F.S. | |
| STATIC Performance | |||
| Static Accuracy | ±0.10...±0.25...±0.50...±1.00 | %FSO BFSL |
|
| Thermal Error | ±0.05............................±0.10 | %FSO/°C | worst case |
| Resolution | Infinitesimal | ||
| Environmental | |||
| Wetted Materials | 316 SS, Fluorocarbon | options available | |
| Compensated Temp Range | 0 to 50........................10 to 30 | °C | options available |
| Operating Temp Range | -10 to 60 | °C | options available |
| Electrical | |||
| Excitation | 2.5 - 10 5 9 - 30 |
VDC | mVdc output (ratiometric) mVdc output (non-ratiometric) mAdc, Vdc output |
| Input Current | 3.5 20 |
mA max | mVdc, Vdc output mAdc output |
| Output | 2.5 - 10 0 - 100 0 - 5 4 - 20 |
mV/V mV VDC mA |
ratiometric, depending on ranges non-ratiometric 3 wire 2 wire |
| Zero Offset | ±5 ±60 ±0.12 |
mV mV mA |
mVdc output Vdc output mAdc output |
| Output Impedance | < 10 | ohms | |
| Insulation Resistamce | 100 | megaohms | at 50 VDC |
| Circuit Protection | Polarity, surge/shorted output | ||
| Physical | |||
| Dimensions | 1.0" dia. X 4.02' long. | ||
| Weight | 198 | grams | excluding cable |
| Cable | Polyurethane jacketed shielded cable with polyethylene vent tube. 90 kilograms pull strength. Conductors are 22 AWG. Tefzel jacket optional. | 70 g/m | specifiy cable length as separate line item |
| Mounting Provisions | Suspended by cable. For turbulent conditions, specify optional mount bracket or conduit fitting | ||