GENEQ > Biotechnology > Pump Syringes
Higher Force Pump
Model KDS 410
| Higher
Force Pump for Demanding Applications |
| BENEFITS |
| • High
pressure dispensing • Automatic dispensing of small volumes • Consistent delivery of fluids • Hands free operation • Accurate fluid delivery |
| FEATURES |
| Quick set up and installation |
| High pressure syringe clamps |
| Typical Accuracy 1% or better |
| Up to 100 lbs linear force |
| Wide variety of syringes from 10 µl to 140 ml |
| Wide plunger travel |
| Quick fluid filling |
| Power reduction mode |
| Volume dispense |
| Minimum flow rate of 0.001 µl/hr using 10 µl syringe |
| Maximum flow rate of 146.7 ml/min. using 140 ml syringe |
| Power failure indication |
| Anti-Syphon |
| TTL interface for remote activation |
| RS-232 |
| Daisy chain up to 99 pumps together |
| APPLICATIONS |
| Continuous delivery of fluid |
| Pilot plant reactor dosing |
| Dispensing Viscous Fluids |
| MARKETS |
| Pharmaceutical |
| Biotech |
| Chemical |
| Petrochemical |
| Neuroscience |
| Research and Devlopment |
| Government |
| Food and Beverage |
| FLOW RATES | |||
Syringes |
Diameter |
Minimum |
Maximum |
10 µl |
0.46 mm |
0.001 µl/hr |
21.05 µl/min |
25 µl |
0.73 mm |
0.003 µl/hr |
53.04 µl/min |
50 µl |
1.03 mm |
0.005 µl/hr |
105.5 µl/min |
100 µl |
1.46 mm |
0.009 µl/hr |
212.1 µl/min |
250 µl |
2.30 mm |
0.021 µl/hr |
526.4 µl/min |
500 µl |
3.26 mm |
0.042 µl/hr |
1.057 ml/min |
1 ml |
4.61 mm |
0.083 µl/hr |
2.115 ml/min |
3 ml |
8.59 mm |
0.287 µl/hr |
7.343 ml/min |
5 ml |
10.30 mm |
0.413 µl/hr |
10.55 ml/min |
10 ml |
14.57 mm |
0.826
µl/hr |
21.12
ml/min |
20 ml |
19.05 mm |
1.411
µl/hr |
36.11
ml/min |
30 ml |
21.59 mm |
1.813
µl/hr |
46.39
ml/min |
60 ml |
26.60 mm |
2.751
µl/hr |
83.12
ml/min |
100 ml |
34.90 mm |
4.736
µl/hr |
121.2
ml/min |
140 ml |
38.40 mm |
5.733
µl/hr |
146.7
ml/min |
| SPECIFICATIONS | |
| Model | Model KDS 410 |
| Syringe Size | 10 µl to 140 ml |
| Fuse | 115 V, 5 x 20 mm, 250 V 0.25 A T slow blow / 240 V, 0.125 A T slow blow |
| Voltage Operating range | US 110 - 125 VAC, 0.25 A, 50/60 Hz / CE 220 - 260 VAC, 0.125 A, 50/60 Hz |
| Drive mechanism | Microprocessor controlled stepper motor 1/2 - 1/16 microstepping, driving a lead screw through a belt and pulley drive mechanism |
| Force | >100 lbs |
| Pusher Advance / Per Micro Step | 1/16 step, 0.165365 micron or 0.0000065 in |
| Volume Per Micro Step | 1/16 step with 60 ml BD syringe 0.0917 µl |
| Speed Range | 2.56 x 104:1 |
| Stepping Rate | |
| Minimum | 1 µ step/120 second |
| Maximum | 1600 1/2 steps/second |
| Linear Travel rate | |
| Minimum | 4.95 x 10 4 cm/hr |
| Maximum | 12.67 cm/hr |
| Flow rate Range | 0.826 µl/hr to 21.12 ml/min using 10 ml syringe |
| Dimensions: H x W x D | 15 x 28 x 24 cm (6 x 11 x 9.5 in) |
| Weight | 6.4 kg (14 lbs) |
| Syringes NOT included | |
| OPTIONS |
| • Alarm Indication at End