| Senseur photométrique |
| Modèle LI-210SA |
|
![]() |
| Description |
The LI-210SA Photometric Sensor utilizes
a filtered silicon photodiode to provide
a spectral response that matches the
CIE curve within ± 5% with most light
sources. This photodiode and filter combination
is placed within a fully cosinecorrected
sensor head to provide the proper
response to radiation at various angles
of incidence.
Some of the applications for the
LI-210SA Photometric Sensor include
interior and industrial lighting, outdoor
illuminance, passive solar energy, architecture
and lighting models, illumination
engineering, and biological sciences that
require illuminance measurements. The
LI-210SA is a research grade photometric
sensor that is very reasonably priced. |
| Specification |
Absolute Calibration: ± 5% traceable to NIST. Sensitivity: Typically 30 µA per 100 klux. Linearity: Maximum deviation of 1% up to 100 klux. Stability: < ± 2% change over a 1 year period. Response Time: 10 µS. Temperature Dependence: ± 0.15% per °C maximum. Cosine Correction: Cosine corrected up to 80° angle of incidence. Azimuth: < ± 1% error over 360° at 45° elevation. Tilt: No error induced from orientation. Operating Temperature: -20 to 65 °C. Relative Humidity: 0 to 100%. Detector: High stability silicon photovoltaic detector (blue enhanced). Sensor Housing: Weatherproof anodized aluminum case with acrylic diffuser and stainless steel hardware. Size: 2.38 Dia.x 2.54 cm H (0.94”x 1.0”). Weight: 28 g (1 oz.). Cable Length: 3.0 m (10 ft.) standard. |
| Order Information |
The LI-210SA Photometric Sensor cable
terminates with a BNC connector that
connects directly to the LI-250 Light Meter
or LI-1400 DataLogger. The 2290 Millivolt
Adapter should be ordered if the LI-210SA
will be used with a strip chart recorder or
datalogger that measures millivolts. The
2290 uses a 604 Ohm precision resistor to
convert the LI-210SA output from
microamps to millivolts. The Photometric
Sensor can also be ordered with bare leads
(without the connector) and is designated
LI-210SZ. The 2003S Mounting and
Leveling Fixture is recommended for each
sensor unless other provisions for mounting
are made. Other accessories are described
on the Accessory Sheet. |
|
| LI-210SA | Photometric Sensor (with BNC connector) |
| LI-210SZ | Quantum Sensor (with bare leads) |
| 2003S | Mounting and Leveling Fixture |
| 2222SB-50 | Extension Cable (50 ft.) |
| 2222SB-100 | Extension Cable (100 ft.) |
| 2290 | Millivolt Adapter |