Microprocessor Based Turbidity Meter
Model ANALITE 160

  
The ANALITE 160 turbidity meter is the latest of our range of turbidity measurement instruments. It is designed to operate with either an ANALITE retro-scatter (180°) probe or an ANALITE 90° probe - to be specified at time of order. The ANALITE 160 turbidity meter allows the user to set up measurement parameters through a user friendly menu system displayed on the in-built 2 line alphanumeric display.

Measurements can be read directly from the display at any time or downloaded to a computer/printer through the optional RS232 output at user selectable periodic intervals. The NEP160 will power up automatically to its last settings whenever external power is applied making it ideal for logging applications when using the analogue output or optional RS232 port.

SPECIFICATIONS
Range 0.0 to20,000NTU (4,000NTU limit on 90° probe) (over four ranges automatically determined).
Display 2 line, 16-character dot matrix alphanumeric liquid crystal display.
Display language English standard, other languages available to order.
Parameters displayed
  • Turbidity (NTU) - default
  • Relative Turbidity Reference (NTU)
  • Relative Turbifity (Turbidity - Relative Turbidity Reference NTU)
  • Date/Time - default
  • Reading updated every 1 second.
    Averaging period 0.5 second or 10 seconds nominal - user selectable.
    Range Steps
  • <0.1 to 20NTU
  • <1 to 200NTU
  • <10 to 2,000NTU
  • <100 to 20,000NTU (using 180° probe only)
  • Resolution
  • 0.02NTU
  • 0.1NTU
  • 1NTU
  • 10NTU
  • Repeatability 2% ± digit on all ranges.
    Data Logging User set for one reading every 1 to 90 seconds or minutes. All readings stored in the Notepad.
    Notepad 100 readings each with time and date.
    Setup procedures
    Menu driven, including:
  • Calibration
  • Automatic Logging
  • Analogue output range selection
  • Reference Turbidity value
  • Setting date and time.
  • Setup memory Non-volatile EEPROM.
    Clock Calendar clock displays date and time. Year 2000 compliant.
    Good Laboratory Practice All readings as well as calibration constants are stored together with the Time and Date and can be recalled at and time.
    Analogue Output 0 - 2 volts full scale corresponding to preset measurement range. Output impedance 600 ohms nominal.
    Power
  • Internal: 6V NiMH rechargeable battery.
  • External: 12 to 16V dc, 100mA max. incl. NiMH Charge current.
  • Power management Automatic power down when operating from batteries after approx. 5 minutes may be selected. Automatic power up when powered externally.
    Low battery indication prior to shut down.
    Measurement outputs Inbuilt LCD, analogue output and optional RS232 port.
    RS232 Port (optional) The optional RS232 port can output readings on request or at preset intervals of time from 1 to 90 seconds or minutes. The Notepad memory can also be downloaded on request. 9600 baud rate, 8 bits, no parity, 1 stop bit, Xon/Xoff protocol.
    Dimensions 187mm x 110 mm x 51 mm (display unit)
    Weight Display unit: 0.5kg
    180° probe: 0.2kg with 2m lead
    90° probe: 0.2kg with 2m lead
    Operating temperature 0 to 50°C
    Operating humidity 0 to 90% RH
    Storage temperature -10 to 60°C
    Case rating IP65

    CAT. NO. DESCRIPTION
    NEP160 Display unit
    NEP160-1 NEP160 with general purpose NEP352 - 180° probe
    NEP160-3 NEP160 with NEP290 - 90° probe