| The MicroVol 1100 low volume air sampler provides a
flexible sampling platform for PM10, PM2.5 or TSP
particulates and basic meteorological parameters. The MicroVol 1100 is suitable for both indoor and outdoor applications. The unit is microprocessor controlled and uses a mass flow sensor in conjunction with ambient temperature and pressure sensors to automatically maintain a constant volumetric flow rate. Particulates are size selected and collected on a sample filter for gravimetric analysis. |
| Standards |
• PM10: AS/NZS 3580.9.9 2006 |
| Indoor sampling |
• Low power consumption |
| Outdoor sampling |
• Wind direction and speed used to activate/de-activate sampler |
| Enhanced communication |
• RS232 output for data collection and networking |
| Specifications |
Volumetric flow range/accuracy: |
1.0 - 4.5 l/min |
Flow accuracy: |
± 2% of reading |
Flow repeatability: |
± 0.5% of reading. |
Temperature range/accuracy: |
0 - 45 deg C ± 1 deg. |
Barometric pressure range: |
600 - 900 Torr ± 4 Torr. |
Filter types: |
47 mm ringed circular filter. |
Inlets available: |
PM10, TSP (standard), PM2.5 (optional) |
Sampler dimensions: |
11.8 in (height) x 6.7 in (diameter)(300 x 170 mm) |
Sampler weight: |
8.3 lbs (3.75 kg) |
Battery pack dimensions: |
7.3 in (height) x 6.7 in (diameter)(185 x 170 mm) |
Battery pack weight: |
8.8 lbs (4.4 kg) |
Battery pack life: |
48 hours sampling from fully charged battery pack. |
Operating voltage: |
12 V DC. |
Power consumption: |
2.5 - 3 watts depending on filter loading. |
Standard Accessories: |
TSP/PM10 size selective inlet. Single 47mm filter holder. 100 - 240V AC to 12V DC power converter. MicroVol Downloader software & RS232 cable. |
| Communications/Data logging |
Number of readings: |
150 averaging periods for each parameter monitored. |
| Averaging periods: | User selectable. |
| External inputs: | 1x wind direction sensor input (10 Kohm potentiometer). 1x wind speed sensor or tipping bucket rain gauge input (contact closure). |
| Data collection: | Via RS232 port. |
| Options |
• Purpose built battery pack, or solar panel and battery pack. |