| SPECIFICATIONS |
| Plate type |
96-well plates |
| Light source |
Tungsten halogen bulb |
| Photodetector |
8 photodiodes, one reference |
| Wavelength range |
380 to 900 nm |
| Dynamic range |
0.000 to 4.000 Abs |
| Accuracy |
±1% ±0.010 Abs, 0.000-2.500 Abs |
| Linearity |
- ±1% from 0.000-2.000 Abs
- ±2% from 2.500-3.500 Abs
|
| Repeatability |
- ±0.5% ±0.005Abs, 0.000-2.500 Abs
- ±1.5% ±0.005Abs, 2.500-3.500 Abs
- ±2.5% ±0.005Abs, 3.500-4.000 Abs
|
| Read method |
Endpoint, kinetic |
| Reading speeds |
- 5 seconds kinetic single well
- 12 seconds, single well
- 20 seconds, single well, dual wavelength
|
| Temperature ctrl |
From 4 ºC above ambient to 50 ºC (I opt.), well-to-well
uniformity ±0.5 ºC |
| Shaking |
User-programmable speeds and timing |
| Diagnostic |
Self-test, test plate software on board |
| Display |
2 x 24 LCD display |
| Keypad |
Membrane keypad w. 25 keys |
| I/O |
RS232 bidirectional serial interface |
| Parallel |
Centronics printer interface |
| Weight |
13.6 kg (30 lb) |
| Dimensions |
40.6 x 39.4 x 22.2 cm (16 D x 15.5 W x 8.75 H in.) |
| Power |
0-240 V @ 50/60 Hz (external transformer) |
| Filters included |
- 405, 450, 490, 630
- 405, 450, 490, 630, 340 (UV option)
|
| Regulatory |
- CE Mark, ETL Mark for UL3101-1
- CAN/CSA C22.2 No.1010.1
- Conforms to IEC 1010-1
|