GENEQ > Biotechnologie > Cryoscopes
| Cryoscope à échantillons multiple |
| modèle Advanced™ 4C3 |
The Advanced 4C3 Multi-Sample
Cryoscope uses the preferred freezing-point method to precisely determine
the amount of added water in milk. The Advanced 4C3 offers unique continuous
through-put for maximum efficiency in large-volume testing. Batch testing
is also efficient with the 4C3; samples can be prepared in advance and
loaded into sample cassettes. |
When ready to begin testing, an operator
can load up to four sample cassettes (holding up to 40 samples) into cryoscope,
press Start and walk away. The instrument automatically tests each sample
in turn and records the results; keeping your cryoscope and operator working
at peak efficiency. |
| FEATURES |
|
| OTHER FEATURES |
| Superior Results |
The Advanced 4C3
freezing-point technology complies with both the International reference
and screening metho; delivering precise, accurate, repeatable and reliable
results |
| Positive Sample Identification |
Availble printer
and RS232 ports from which, date, time, sample ID and test result can
be reported |
| Operator-selectable Options |
Standard design
includes m°C or m°H scales, timed plateau or plateau seeking modes,
and four language prompts. |
| Calibration Ease |
Simply load the
calibration samples into the specially coded calibration cassette and
calibration is automatically triggered to run unattended |
| Fast Testing, Fully Automated |
With the unique
dual-stage, solid-state cooling assembly, testing can begin immediately.
You don't waste time waitin for your instrument to cool down. |
| Rewarding |
The capacity of
the 4C3 allows your lab to efficiently test and manage supplied milk in
order to properly compensate producers, ensure quality products and profitable
returns. |
| SPECIFICATIONS | |
| Sample Size | 2.0 to 2.5mL |
| Sample Capacity | Holds up to 4 (four) cassettes |
| Readout | 20-character digital display and on-board thermal printer |
| Units | m°C or m°H |
| Range | 0 to -1000 m°C or m°H |
| Linearity | Typically within ±0 5% (straight line method) |
| Resolution | ±1 m°C or m°H |
| Repeatability | ±2 m°C (1 S.D.) |
| Power Requirement | 100 to 130 V(60Hz) or 200 to 250 V(50 Hz) |
| Power Consumption | 350 W |
| Dimensions | 21.5"H x 21.5"W x 20"D (55cmH x 55cmW x 51cmD) |
| Net Weight | 55 lbs. (25 kg) |
| Shipping Weight | 75 lbs. (34 kg) |
| Storage Temperature | -40 to +158°F (-40 to +70°C) |
| Operating Temperature | 64 to 95°F (18 to 35°C) |
| Drift | Less than 1 digit per month plus 1 digit for every 9°F or 5°C ambient temperature change |
| Room Humidity | 5 to 80% R.H., non-condensing |
| Sample Throughput | Approximately 40 samples per hour in the 30-second plateau mode at an ambient temperature of 23°C. Increase in ambient temperature will result in decrease of sample throughput. |
| Warranty | One-year limited warranty on workmanship and all parts except glass, plastic, and parts warranted by their makers. |