GENEQ > Environnement > Air - Échantillonneurs > Compteurs de Particules

Échantillonneur et compteur de particules
Modèle SidePak AM510


The SidePak AM510 personal aerosol monitor for industrial hygiene and safety professionals displays and data logs aerosol concentration in real time. The rugged, belt-mountable laser photometer is compact and quiet, minimizing interference and worker discomfort. The built-in sampling pump lets you attach a wide variety of size-selective aerosol inlet conditioners for breathing zone or area measurements with a respirable cyclone or one of the three integrated impactors.

The SidePak monitor incorporates TSI's Smart Battery Management System. This system utilizes a built-in "gas gauge" in the battery packs to monitor battery condition and provide precise run time information.

The SidePal monitor's easy-to-read display shows your data as both real-time aerosol mass-concentration and 8-hour time-weighted average (TWA). The monitor's intrinsic safety rating makes it ideal for hazardous/explosive environments. With its convenient data logging and long battery life, the SidePak is also ideal for extended sampling. The easy-to-use TrakPro Data Analysis Software lets you create effective graphs and reports.


USER FRIENDLY
  • Small, lightweight and quiet to maximize worker acceptance
  • Rugged design with secure belt clip
  • Easy-to-understand user interference with only four keys
  • Lockable keypad prevents tampering while sampling
  • User-adjustable sample flow rate
  • Define, lable and store multiple calibration constants
  • Easy-to-read LCD display
  • Convenient, threaded tripod socket accomodates area sampling

ADVANCED FEATURES
  • Smart battery management system provides precise run time information, maximizes battery capacity and speeds charging
  • Integrated pump allows use of size-selective aerosol inlet conditioners
  • Built-in impactors let you choose "non" 1.0, 2.5 or 10 micron cut off
  • Display shows real-time concentrations (mg/m3) and "on the fly" TWA as you data log
  • Display statistic: max, min and average readings, elapsed time and 8-hour TWA

QUICK AND EASY REPORTS
  • Convenient preprogramming for occupational exposure sampling
  • Data log for long periods and store multiple tests
  • Analyze data, print graphs and create reports with TrakPro software
  • USB port lets you conveniently connect to your computer

POWER TO SPARE
  • Long-lasting NiMH rechargeable battery packs eliminate "memory" issues
  • Choice of rechargeable NiMH smart battery packs or AA-cell pack

SPECIFICATIONS
Sensitivity
Sensor Type 90° light scattering, 670 nm laser diode
Aerosol Concentration Range 0.001 to 20mg/m3 (calibrated to respirable fraction of ISO 12103-1, A1 test dust)
Particle Size Range 0.1 to 10 micrometer (µm)
Minimum Resolution 0,001 mg/m3
Zero Stability ±0.001 mg/m3 over 24 hours using 10-second time-constant
Temperature Coefficient approximately +0.0005 mg/m 3 per °C (for variations from temperature at which instrument was last zeroed)
Flow Range
Range User adjustable 0.7 to 1.8 liters/min (lpm)
Temperature Range
Operating Range 0 to 50°C (32 to 120°F)
Storage Rage -20 to 60°C (-4 to 140°F)
Time Constant (LCD display)
Data Points Approx. 31000 (21 days logging once per minute)
Logging Interval User adjustable, 1 second to 1 hour
User-Select Calibration Factors
Factory Setting 1.0 (non-adjustable)
User-defined Settings 3, with user-defined lables
Range 0.1 to 10.0, user-adjustable
Physical
External Dimensions
  • 106 x 92 x 70 mm (4.2 x 3.7 x 2.8 in.) - with 801724 or 801729 batteries
  • 130 x 92 x 70 mm (5.1 x 3.7 x 2.8 in.) - with 801708, 801722, 801728 or 801736 batteries
Weight
  • 0.46kg (16oz) with 801724 or 801729 batteries
  • 0.54kg (19oz) with 801708, 801722, 801728 or 801736 batteries
Display 2 line x 12 character LCD
Tripod Socket 1/4 - 20 female thread
Power Supply / Charger (P/N 2613210)
Input Voltage Range 100 to 240VAC, 50/60Hz
Output Voltage 9 VDC @ 1.0 A
Battery Charge Time 6.5 hours
Maintenance
Factory Clean/Calibrate Recommended annually
User Zero Calibration Before each use
User Flow Calibration As needed
Communications Interface
Type USB 1.1
Connector, Instrument USB Mini-B (soket)
Minimum Computer Requirements for TrakPro software
Communications Port Universal Serial Bus (USB) v1.1 or higher
Operating System Microsoft Windows ® 98, Me, 2000, XP (USB not supported in Windows 95 or NT)
Approvals
Approvals TSI Battery Pack P/Ns: 801722 or 801724
Intrinsic Safety INTRINSICALLY SAFE CSA
Rating Information Exia
Class 1 Groups A, B, C, D
T2A with 801724
T2C with 801722
File: 200507
TSI Battery Pack P/N: 801708
NOT INTRINSICALLY SAFE

BATTERY INFORMATION
Battery Option
Charge Time*
Intrinsically Safe
1650 mAH NiMH Pack (P/N 801724)
3.5 Hours
Yes**
2700 mAH NiMH Pack (P/N 801722)
5.5 Hours
Yes**
6-Cell AA-size Pack (P/N 801708)
NA
No
* Of a fully deleted battery
** All dust plugs and gaskets must be installed

TYPICAL BATTERY LIFE
Battery Pack
Cell Type
Hours @ 0.7 lpm
Hours @ 1.7 lpm
1650 mAH NiMH Pack (P/N 801724)
NiMH (4.8 V, 1650 mAH)
11.5
9.2
2700 mAH NiMH Pack (P/N 801722)
NiMH (4.8 V, 2700 mAH)
19.8
15.6
6-Cell AA-size Pack (P/N 801708, with six user-supplied AA cells)
AA Alkaline Cells
29.6
22.5
AA NiMH Cells (1600 mAH per cell. Cells cannot be charged while inside AA Pack)
8.8
6.7