GENEQ > Environment > Pumps
Pump Calibrators
Models 4046 & 4146
![]() |
Personal
sampling pumps are key tools for exposure assesment. The accuracy of sampling
depends on how much air passes through the sampling media and good laboratory
analysis. As a common practice, sampling pumps should be calibrated before
and rechecked after sampling using a calibration device and procedure. TSI
offers two configurations for calibrating pumps that draw an active air
sample.
The Model 4146 pump calibrator offers a range of 0.01 to 20 lpm and Model
4046, from 2.5 to 300 lpm. These small, portable, battery-operated units
are lightweight ad easy to se. Volumetric flowrate is displayed continuously
so adjustments to pump flow can be made in real-time. The flowmeters are
temperature-compensated and pressure-corrected to calculate volumetric flow
rate at an accuracy within 2% of reading.
| COMES COMPLETE WITH |
|
| SPECIFICATIONS | ||
|
Model 4146
|
Model 4046
|
|
| Flow Measurement Range |
0.1 to 20 lpm
|
2.5 to 300 lpm
|
| Accuracy |
±2% or 0.005 std. lpm, which ever is greater
|
±2% or 0.05 std. lpm, which ever is greater
|
| Dimensions |
5 x 2 x 1.25 in.
(12.7 x 5 x 3.2 cm) |
7.2 x 2.5 x 2.1 in.
(18.3 x 6.4 x 5.3 cm) |
| Shipping Weight |
2.2 lbs (1.0 kg)
|
3.0 lbs (1.1kg)
|
| Warm-up Time |
1 min.
|
1 min.
|
| DC Power Input |
7.5 VDC, ±1.5V, 300 mA max
|
7.5 VDC, ±1.5V, 300 mA max
|
| CAT. NO. | DESCRIPTION |
| TSI4146-00 | Pump Calibrators Model 4146 |
| TSI4046-00 | Pump Calibrators Model 4046 |