GENEQ > Biotechnology > pH Meters
Easy-to-Operate Precision pH Meters
Models pH 210/211/211R
Combining the most frequently asked for parameters, the pH 211 and pH 211R are the popular choice for laboratory bench top instruments around the world. You can measure pH directly, as well as milliVolts for pH, Ion Specific Electrodes (ISE), and Oxidation Reduction Potential (ORP) electrodes. For added versatility pH 211R offers analog output of 0 to 5 V that represents the full pH and mV range. Temperature compensation, either automatic or manual, is your choice and easy to do. The instructions for calibration are presented in a step-by-step display with safeguards to prevent calibration errors.
When your need is solely to measure pH, the easy
choice is pH 210. Calibration is just a few simple steps and is guided by graphics
on the LCD. You can calibrate at 1 or 2 points by choosing any combination of
the 5 buffer values in the meter's powerful memory. These meters offer the best
value in bench top pH measurement today!
| SPECIFICATIONS | |||
|
pH210
|
pH211
|
pH211R
|
|
| Range (pH) |
0.00 to 14.00
|
0.00 to 14.00
|
|
| Range (mV) |
-
|
±399.9 (ISE); ±1999 (ORP)
|
|
| Range (°C) |
0.0 to 100.0
|
0.0 to 100.0°C
|
|
| Resolution (pH) |
0.01
|
0.01
|
|
| Resolution (mV) |
-
|
0.1 (ISE); 1 (ORP)
|
|
| Resolution (°C) |
0.1
|
0.1
|
|
| Accuracy (pH) (@20°C/68°F) |
±0.01
|
±0.01
|
|
| Accuracy (mV) (@20°C/68°F) |
-
|
±0.2 (ISE); ±1 (ORP)
|
|
| Accuracy (°C) (@20°C/68°F) |
±0.5
|
±0.5
|
|
| Typical EMC Deviation (pH) |
±0.03
|
±0.03
|
|
| Typical EMC Deviation (mV) |
-
|
±2
|
|
| Typical EMC Deviation (°C) |
±0.3
|
±0.3
|
|
| pH Calibration |
Automatic 1 or 2 points with 5 memorized
buffers (pH 4.01, 6.86, 7.01, 9.18, 10.01)
|
||
| Temperature Compensation |
Automatic (with HI 7669/2W probe) or
manual from 0 to 100°C (32 to 212°F)
|
||
| Electrode |
HI 1131B pH electrode, combination, glass,
interchangeable, BNC connector, 1 m (3.3) cable (included)
|
||
| Temperature Probe |
HI 7669/2W with 1 m (3.3) cable
(included)
|
||
| Input Impedance |
1012 Ohm
|
||
| Analog Output |
-
|
-
|
0 to 5 V (pH 211R only)
|
| Power Supply |
12VDC (included)
|
||
| Environment |
0 to 50°C (32 to 122°F); RH 95%
|
||
| Dimensions |
240 x 182 x 74 mm (9.4 x 7.2 x 2.9)
|
||
| Weight |
1.1 Kg (2.4 lb.)
|
||
| CAT. NO. | DESCRIPTION |
| pH 210 | Supplied complete with HI 1131B interchangeable pH electrode, BNC, 1 m (3.3) cable, HI7669/2W temperature probe, pH 4 and 7 calibration solutions, electrolyte solution, 12VDC adapter and instruction manual. |
| pH 211 | Supplied complete with HI 1131B interchangeable pH electrode, BNC, 1 m (3.3) cable, HI7669/2W temperature probe, pH 4 and 7 calibration solutions, electrolyte solution, 12VDC adapter and instruction manual. |
| pH 211R | Supplied complete with HI 1131B interchangeable pH electrode, BNC, 1 m (3.3) cable, HI7669/2W temperature probe, pH 4 and 7 calibration solutions, electrolyte solution, 12VDC adapter and instruction manual. |
| HI 1131B | Combination pH electrode |
| HI 76404 | Electrode holder |
| HI 7004L | pH 4.01 buffer solution, 500 mL |
| HI 7007L | pH 7.01 buffer solution, 500 mL |
| HI 7010L | pH 10.01 buffer solution, 500 mL |
| HI 77400P | Cal. kit (pH 4 & 7, 20 mL, 5 pcs each) |
| HI 70300L | Storage solution, 460 mL |
| HI 8427 | pH and ORP electrode simulator |
| HI 931001 | pH and ORP electrode simulator |