GENEQ > Environnement > Air Échantillonneurs

Pompe d'échantillonnage personnelle,
Modèle SidePak™ SP330 & SP350


The SIDEPAK SP330 and SP350 pumps are rugged, basic pumps. They offer a choice of battery packs that incorporate TSI’s Smart Battery Management System. The SP330 is a 3-liter pump, and the SP350 is a 5-liter pump.

FEATURES

Smart Battery Management System

  • Minutes of run time are displayed
  • Real-time battery condition, not just an estimate of % battery remaining
  • Integrated microchip technology

Multiple Battery Options

  • Rechargeable 1650 mAH or 2700 mAH NiMH battery packs
  • 6-cell AA-size battery pack

User keypad programming

  • Sample run time
  • Flow setpoint
  • Delayed start
  • Keypad lock

SPECIFICATIONS
 
SP330
SP350
Flow Range
Constant Flow
500 to 3,000 cc/min
500 to 4,000 cc/min
Constant Pressure
20 to 800 cc/min (requires low flow adapter)
Flow Control
Back EMF
±5% of constant flow setpoint
Flow Performance
Flow vs. Max Vacuum Pressure
  • 500 cc/min @ NA in. H2O
  • 1,000 cc/min @ 55 in. H2O
  • 1,500 cc/min @ 40 in. H2O
  • 2,000 cc/min @ 28 in. H2O
  • 2,500 cc/min @ 19 in. H2O
  • 3,000 cc/min @ 10 in. H2O
  • 3,500 cc/min @ 0 in. H2O
  • 500 cc/min @ NA in. H2O
  • 1,000 cc/min @ 32 in. H2O
  • 1,500 cc/min @ 32 in. H2O
  • 2,000 cc/min @ 29 in. H2O
  • 2,500 cc/min @ 25 in. H2O
  • 3,000 cc/min @ 19 in. H2O
  • 3,500 cc/min @ 13 in. H2O
  • 4,000 cc/min @ 7 in. H2O
  • 4,500 cc/min @ 0 in. H2O
Flow Fault
If flow is blocked and drops by more than 3% for more than 20 seconds, the pump stops and retains all sample data. Auto-restart is attempted after 40 seconds up to 10 times.
Flow Indicator
Adjust Flow Menu Screen, range 0.0 to 99.9% for both
Inlet Pulsation Ratio
8%@ 2.0 lpm w/GF filter
9%@ 2.0 lpm w/GF filter
User Adjustable
Sample Run Time
Values: Flow Setpoint, Delayed Start, Key Pad Lock
Recorded Values Temperature Range: Total Sample Time
Operating Range
0 to 45°C (32 to 112°F)
Storage Range
-20 to 60°C (-4 to 140°F)
Physical
External Dimensions
With 801724 battery: 106 mm x 92 mm x 70 mm (4.2 in. x3.7 in. x2.8 in.)
With 801722 or 801708 battery: 130 mm x 92 mm x 70 mm
(5.1 in. x 3.7 in. x 2.8 in.)
Weight
  • 0.46 kg (16 oz) with 801724 battery
  • 0.54 kg (19 oz) with 801722 battery
  • 0.54 kg (19 oz) with 801708 battery
Display
2 line x 12 character LCD
Tripod Socket
1/4–20 female thread
Power Supply (P/N 2613210)
Input Voltage Range
100 to 240 VAC, 50 to 60 Hz
Output Voltage
9 VDC @ 1.0 A
Maintenance
Factory clean and test
Recommended annually
User flow calibrations
Before and after each use
Approvals
TSI Battery Pack P/Ns
801722 or 801724
Intrinsic Safety
INTRINSICALLY SAFE CSA
Rating Information
  • Exia
  • Class 1 Groups A, B, C, D
  • T2A with 801724
  • T2C with 801722
  • File: 200507
  • TSI Battery Pack P/N: 801708
    NOT INTRINSICALLY SAFE
Immunity
EN61326-1:1997 + A11998 Clause 6
Emissions
EN61326:1997 + Amendment A1:1998

BATTERY INFORMATION
Battery Option
Charge Time*
Intrinsically Safe
1650 mAH NiMH Pack
(P/N 801724)
3.5 Hours
Yes**
2700 mAH NiMH Pack
(P/N 801722)
5.5 Hours
Yes**
6-Cell AA-size Pack
(P/N 801708)
NA
No
* Of a fully depleted battery
** All dust plugs and dust gaskets must be installed

BATTERY LEVEL INDICATOR
The Smart Battery Management System utilizes a built-in “gauge” in the SIDEPAK battery packs. The gauge monitors battery capacity and calculates run time information by dividing capacity of the battery (mAH) by the instantaneous current consumed by the instrument (mA). This calculation is correct for current operating conditions and can change due to current (mA) consumption or changes in battery capacity.

TYPICAL COLLECTION MEDIA BACK PRESSURES

BATTERY LIFE PERFORMANCE (estimated hours)
*** Using Energizer AA-size, E91 alkaline batteries

 

SOFTWARE, ACCESSORIES & KIT