The SIDEPAK Model SP530 pump offers
all the features of the SP330 plus an
additional precision flowmeter that acts as a secondary flow measurement
standard. Unlike conventional pumps, this 3-liter pump continuously measures
and data logs flowrate over a very wide range to provide precise control,
accurate total sample volume and average flowrate information.
|
| FEATURES |
|
Advanced precision flowmeter control |
|
|
Multi-mode operation |
|
|
Data logging of flow |
|
| TRAKPRO Data Analysis Software |
|
| Unparalleled flowrate calibration |
|
| SPECIFICATIONS | |
| Flow Range | |
| Constant Flow | 500 to 3,000 cc/min |
| Constant Pressure | 20 to 800 cc/min (requires low flow adapter) |
| Flow Control | |
| Hybrid Control | Thin-film flowmeter and back EMF ±3% of constant flow setpoint |
| Thin-Film Flowmeter | |
| Flow Range | 50 to 3,000 cc/min |
| Flow Resolution | 1 cc/min |
| Operating Pressure Range | 12 to 17 psia |
| Operating Temp. Range | 0 to 45°C (32 to 112°F) |
| Flow Performance | |
| Flow vs. Max. Vacuum Pressure |
|
| Flow Fault | If flow is blocked and drops by more than 3% for more than 20 seconds, the pump stops and retains all sample data. Auto-restart is attempted after 40 seconds up to 10 times. |
| Flow Indicator LCD | Adjust Flow Menu Screen, range 500 to 3,000 cc/min |
| Inlet Pulsation Ratio | 4% @ 2.0 lpm w/GF filter |
| Data Logging | |
| Data Points | Approx. 31,000 data points (21 days logging once per minute) |
| Logging Interval | Non-adjustable, 1-minute only |
| User Adjustable | |
| Values |
|
| Recorded Values |
|
| Temperature Range | |
| Operating Range | 0 to 45°C (32 to 112°F) |
| Storage Range | -20 to 60°C (-4 to 140°F) |
| Physical | |
| External Dimensions |
|
| Weight |
|
| Display | 2 line x 12 character LCD |
| Tripod Socket | 1/420 female thread |
| Power Supply (P/N 2613210) | |
| Input Voltage Range | 100 to 240 VAC, 50 to 60 Hz |
| Output Voltage | 9 VDC @ 1.0 A |
| Software/Hardware Specifications | |
| Communications Interface Type | USB 1.1 |
| Connector, Instrument | USB Mini-B (socket) |
| Min. Computer Requirements for TRAKPRO Software | |
| Communications Port | Universal Serial Bus (USB) v 1.1 or higher |
| Operating System | Microsoft Windows® 98, ME, 2000, XP (USB not supported in Windows® 95 or NT) |
| Maintenance | |
| Factory clean and test | Recommended annually |
| User flow calibrations | Before and after each use |
| Approvals | |
| TSI Battery Pack P/Ns | 801722 or 801724 |
| Intrinsic Safety | INTRINSICALLY SAFE CSA |
| Rating Information | Exia Class 1 Groups A, B, C, D T2A with 801724 T2C with 801722 File: 200507 TSI Battery Pack P/N: 801708 NOT INTRINSICALLY SAFE |
| Immunity | EN61326-1:1997 + A11998 Clause 6 |
| Emissions | EN61326:1997 + Amendment A1:1998 |
| BATTERY INFORMATION | ||
|
|
Charge Time*
|
Intrinsically Safe
|
| 1650 mAH NiMH Pack (P/N 801724) |
3.5 Hours
|
Yes**
|
| 2700 mAH NiMH Pack (P/N 801722) |
5.5 Hours
|
Yes**
|
| 6-Cell AA-size Pack (P/N 801708) |
NA
|
No
|
| * Of a fully depleted battery | ||
| ** All dust plugs and dust gaskets must be installed | ||
| BATTERY LEVEL INDICATOR |
The Smart Battery Management System
utilizes a built-in gauge in the SIDEPAK battery packs.
The gauge monitors battery capacity and calculates run time information
by dividing capacity of the battery (mAH) by the instantaneous current
consumed by the instrument (mA). This calculation is correct for current
operating conditions and can change due to current (mA) consumption
or changes in battery capacity. |
| TYPICAL COLLECTION MEDIA BACK PRESSURES |
![]() |
| BATTERY LIFE PERFORMANCE (estimated hours) |
![]() |
| *** Using Energizer AA-size, E91 alkaline batteries |