| SPECIFICATIONS |
| Power Requirements |
115 +/- 20 VAC
230 +/- 40 VAC
20 VA |
| Line Frequency |
45 to 65 Hz |
| Stability |
Maximum drift 0.0005 AU per hour after 30 minute warm-up
|
| Rise Time |
10-90% rise time is 0.6 sec on 0.5 AU range |
| Full-Scale Absorbance Ranges |
0.01, 0.02, 0.05, 0.1, 0.2, 0.5, and 1.0 Absorbance
Units |
| Linearity |
Maximum error 1.5% of full scale at 254 nm |
| Noise Level |
less than 0.0003 AU at 254 nm with dry flow paths after
30 minute warm-up |
| Weight |
4.1 kg (9 lb) |
| Operating Temperature |
0°C to 40°C, noncondensing |
| Display Accuracy |
+/- 3 counts |
| CAT. NO. |
DESCRIPTION |
| 142600 |
Model 280 UV Detector, 115V, Domestic US, includes
both 254 and 280 nm |
| 142602 |
Model 280 UV Detector, 230V, European, CE Marked,
include both 254 and 280 nm |
| Accessories |
| 142608 |
254/280 nm Filter slide. (This combination
slide is included with the Spectra/Chrom UV Monitor.) |
| 142642 |
5 mm pathlength dual flow cell. (This flow
cell is included with Spectra/Chrom UV Monitor.) Illuminated volume 34 µL;
hold-up volume 134 µL (per flow path). |
| 142644 |
2 mm pathlength dual flow cell. Illuminated
volume 14 µL; hold-up volume 54 µL (per flow path). |
| 142640 |
10 mm pathlength dual flow cell. Illuminated
volume 68 µL; hold-up volume 268 µL (per flow path). |