GENEQ > Environment > Ozone
| Ozone Monitor Series 200 |
| Specification | ||
| Series 200 Monitor Base | ||
| Measurement units | ppm | |
| Removable / replaceable sensor head | Yes | |
| On-board alarm | No | |
| Alarm status displayed | No | |
| Low battery indication | Yes | |
| Alarm mute function | No | |
| Sensor condition status | Yes | |
| Stand-by mode | Yes | |
| External signal for alarms & control | None | |
| External signal functions | None | |
| Analogue output | None | |
| Power supply | 12 VDC, 800 mA | |
| Rechargeable battery pack (2 x options) | Standard | 9.6V ( 940mA/hr Ni-Cd) |
| Long Life | 9.6V (2100mA/hr Ni-MH ) | |
| Portable | Yes | |
| Permanently fixable | Screw fix | |
| Remote Sensor Capability | Yes | |
| Enclosure rating | IP20 & NEMA 1 equivalent | |
| Size (with sensor head) | 195 x 122 x 54 (mm); 7> x 43/4 x 2@ (in) | |
| Weight (with sensor head and battery) | < 460 g; < 16 oz | |
| Approvals | Part 15 of FCC Rules EN 50082-1: 1997 EN50081-1: 1992 |
|
| Ozone Sensor Head Specifications | |||
| Sensor Heads | Low | High | Leak |
| Measurement range | 0.000 - 0.500 ppm | 0.50 - 20.00 ppm | 0.00 - 50.00 ppm |
| Accuracy | < 0.008
ppm (0 - 0.100 ppm) ± 10% (0.100 0.500 ppm) |
± 10%
(0.20 - 2.00 ppm) ± 15% (2.00 - 20.00 ppm) |
± 20%, 15 seconds after Reset |
| Resolution |
0.001 ppm | 0.01 ppm | 0.01 ppm |
| Controlled Atmosphere Version | Available on Request | Available on Request | Available on Request |
| T90 Response | < 60 seconds (T90) | < 35 seconds (T90) | < 10 seconds |
| Sensor type | Gas-sensitive semiconductor | ||
| Operating temperature range | -5°C to 50°C; 23°F to 122°F (sensor head and base unit) | ||
| Relative humidity limit | 95% maximum (sensor head and base unit) | ||
| VOC Sensor Head Specification* | |||
| Measurement range | 0 to 400 ppm toluene | ||
| Accuracy | < +/- 10 ppm (0
200 ppm) < +/- 10% ( > 200 ppm) |
||
| T90 response | < 60 s | ||
| Sensor type | Gas-sensitive semiconductor | ||
| Operating temperature range | -20°C to 60°C | ||
| Operating relative humidity range | 5 to 95% non-condensating | ||
| * Specific Calibrations to other VOCs available contact Aeroqual Sales | |||
| Ammonia Sensor Head Specification | |||
| Measurement range | 0 to 1000 ppm | ||
| Accuracy | < +/- 5 ppm (0 100 ppm) < +/- 10% (100 1000 ppm) | ||
| T90 response | < 60 s | ||
| Sensor type | Gas-sensitive semiconductor | ||
| Operating temperature range | -20°C to 60°C; | ||
| Operating relative humidity range | 5 to 95% non-condensating | ||
| CO Sensor Head Specification | |||
| Measurement range | 0 to 2000 ppm carbon monoxide | ||
| Accuracy | < +/- 10 ppm in the range 0 to 400 ppm | ||
| T90 response | < 150s | ||
| Sensor type | Gas-sensitive semiconductor | ||
| Operating temperature range | 0°C to 70°C; | ||
| Operating relative humidity range | 5 to 95% non-condensating | ||