Geneq > Météorologie > Enregistreurs – Miniatures

Enregistreur de pression, température et d’humidité relative à 5 canaux
Modèle SmartReader 4


The SmartReader 4 data logger records pressure, temperature and relative humidity. Use it to monitor and troubleshoot pneumatic systems. With a variety of pressure sensors available, you can tailor the logger to suit your needs. As an added bonus, the SmartReader 4 can tailor the logger to suit your needs. As an added bonus, the SmartReader 4 can also monitor temperature and relative humidity. All probes and sensors are sold separately.

Applications
Tracking performance of pneumatically controlled valves; verifying thermostat and humidistats; monitoring sensitive shipments, etc.

General Specifications
Size 107mm x 74mm x 22mm (4.2” x 2.9” x 0.9”)
Weight 110 g (3.75 oz)
Case Material Noryl® Plastic
Battery 3.6 volt Lithium, 1 Amp-Hour
Resolution 8-bit (1 part in 256)
Mounting Magnetic backing or locking eyelet
Clock Accuracy ± 2 seconds per day
Sampling Methods Continuous (First-in First-out), Stop When Full (Fill-then-stop)
Operating Limits -40°C to 70°C (-40°F to 158°F) and 0 to 95% RH (non-condensing)
PC Requirements Windows PC with at least one free USB or serial port (depending on interface)
Software Requirements TrendReader® 2 (Compatible with Windows XP SP3, Windows Vista SP2& Windows 7 (32 bit & 64 bit)
Memory Size 32 KB (capable of storing up to 32,767 readings)
Sampling Rates User selectable rates from 8 seconds to 5 days in 4-second increments
Number of Channels Five (one internal for ambient temperature, one for optional internal RH sensor, one for optional pressure sensor, and two for optional remote RH and tempera ture probe [model EH-020A])

Sensor Specifications
Internal Temperature Sensor
Type NTC Thermistor - 10,000 Ohm @ 25°C (77°F)
Range -40°C to 70°C (-40°F to 158°F)
Accuracy ±0.2°C over the range of 0°C to 70°C (± 0.3°F over the range of 32°F to 158°F)
Resolution 0.4°C (0.7°F) @ 25°C; better than 1°C (1.8°F) between -25°C and 70°C
(-13°F and 158°F); better than 2.0°C (3.6°F) between -40°C and -25°C (-40°F and -13°F)
Pressure Sensors Range Resolution Cat. No.
PM-005-G 0 to 5 psi (30 kPa) Gauge 0.03 psi (0.21 kPa) 01-0045
PM-030-A 0 to 30 psi (200 Kpa) Absolute 0.15 psi (1.0 kPa) 01-0046
PM-030-G 0 to 30 psi (200 kPa) Gauge 0.15 psi (1.0 kPa) 01-0047
PM-100-A 0 to 100 psi (700 kPa) Absolute 0.5 psi (3.5 kPa) 01-0048
PM-100-G 0 to 100 psi (700 kPa) Gauge 0.5 psi (3.5 kPa) 01-0049
PM-150-G 0 to 150 psi (1000 kPa) Gauge 0.5 psi (3.5 kPa) 01-0086
Calibrated Accuracy ±0.5% of full scale @ 25°C (77°F)

Ordering Information
Item Cat No.
SR-004 (SmartReader 4) 01-0027